Ventilator lessons Lesson 7 of 9
High pressure alarm on the ventilator: what to check
Updated
The high pressure alarm means airway pressure reached the limit you set. On most ventilators inspiration stops there, so the breath can be smaller than set. Start with the patient, then use plateau pressure to find where the pressure comes from.
Step 1: the patient
If oxygen saturation or blood pressure is falling, or the chest is not rising, do not troubleshoot the screen. Disconnect, ventilate by hand with oxygen and feel how hard it is. Pass a suction catheter to check that the tube is open, confirm its position, and look for a tension pneumothorax. Many clinicians remember these as DOPE: displacement, obstruction, pneumothorax, equipment.
Step 2: find the cause with peak and plateau
In a stable, passive patient on volume control, add an inspiratory pause and read the plateau. Then compare it with the last value you trust (peak vs plateau pressure; to put numbers on it, calculate static compliance and resistance).
| Model lungs | Ppeak | Pplat | Gap | Pattern |
|---|---|---|---|---|
| Baseline | 21.7 | 12.5 | 9.2 | |
| Resistance ×3 | 40.1 | 13.0 | 27.1 | Gap wide, plateau the same |
| Compliance ×0.5 | 28.2 | 19.0 | 9.2 | Both up, gap the same |
Pressures in cmH₂O. Volume control, tidal volume 450 mL, rate 12/min, PEEP 5 cmH₂O, Ti 1.0 s with a 0.5 s pause.
Gap wide, plateau about the same: the airways
Something between the ventilator and the alveoli is narrower. Look for a kinked or bitten tube, secretions or a mucus plug, bronchospasm, and water or a blocked filter in the circuit. Suction, add a bite block, give a bronchodilator when wheeze points to bronchospasm, and check the circuit.
Plateau up as well: the lungs, the pleura, or trapped air
First do an expiratory hold. If total PEEP is well above the set PEEP, air is trapped (auto-PEEP). In model lungs with bronchospasm, raising the rate from 14 to 22/min raised Ppeak from 24.4 to 32.9 cmH₂O and total PEEP from 8.9 to 12.8 cmH₂O: the fix there is a slower rate, a longer expiration and treating the bronchospasm.
If total PEEP is not raised, the respiratory system is stiffer: mainstem intubation (check tube depth and listen to both sides), pneumothorax, pulmonary edema, worsening ARDS or pneumonia, atelectasis, or a tense abdomen. Examine the patient and use a chest x-ray or ultrasound.
Sudden or gradual?
- Sudden: biting, a kinked tube, a mucus plug, the tube slipping into one bronchus, pneumothorax, bronchospasm, coughing or fighting the ventilator.
- Gradual: secretions building up, worsening edema or ARDS, a rising abdominal pressure, air trapping building over many breaths.
Do not just raise the limit
Raising the alarm limit silences the alarm, not the cause, and lets higher pressure reach the lungs. In pressure control the pressure is set, so a stiffer lung shows up as a smaller tidal volume and a low volume alarm instead (pressure vs volume control).
Key point
Patient first. Then the plateau: a wide gap points to the tube and airways, a high plateau to trapped air or the lungs, pleura, chest wall and abdomen.
Try it
- Open the MV Simulator, choose Advanced, set Tpause to 0.5 s and press Confirm.
- Choose the Airway plug patient and compare Ppeak with Pplat. Then Stiff lungs. Which one moved the plateau?
- Choose Bronchospasm, raise the rate, and press Exp. hold to see the trapped pressure.
References
- Hess DR. Respiratory mechanics in mechanically ventilated patients. Respir Care 2014;59(11):1773-1794. doi:10.4187/respcare.03410
- Marini JJ. Dynamic hyperinflation and auto-positive end-expiratory pressure: lessons learned over 30 years. Am J Respir Crit Care Med 2011;184(7):756-762. doi:10.1164/rccm.201102-0226PP
- Leatherman JW, Ravenscraft SA. Low measured auto-positive end-expiratory pressure during mechanical ventilation of patients with severe asthma: hidden auto-positive end-expiratory pressure. Crit Care Med 1996;24(3):541-546. doi:10.1097/00003246-199603000-00028