Ventilator lessons Lesson 7 of 9

High pressure alarm on the ventilator: what to check

Updated

The high pressure alarm means airway pressure reached the limit you set. On most ventilators inspiration stops there, so the breath can be smaller than set. Start with the patient, then use plateau pressure to find where the pressure comes from.

Flowchart. High pressure alarm, then look at the patient. If unstable, disconnect, ventilate by hand with oxygen, check the tube with a suction catheter and think of tension pneumothorax. If stable, do an inspiratory pause and compare peak with plateau: a wide gap with the plateau unchanged points to resistance; a raised plateau points to trapped air or the lungs, pleura, chest wall or abdomen.
The order to work through. The flowchart is a teaching summary, not a local protocol.
MV SimulatorSort the cause on the modelSwitch between airway plug, stiff lungs and bronchospasm and watch which pressure moves.

Step 1: the patient

If oxygen saturation or blood pressure is falling, or the chest is not rising, do not troubleshoot the screen. Disconnect, ventilate by hand with oxygen and feel how hard it is. Pass a suction catheter to check that the tube is open, confirm its position, and look for a tension pneumothorax. Many clinicians remember these as DOPE: displacement, obstruction, pneumothorax, equipment.

Step 2: find the cause with peak and plateau

In a stable, passive patient on volume control, add an inspiratory pause and read the plateau. Then compare it with the last value you trust (peak vs plateau pressure; to put numbers on it, calculate static compliance and resistance).

Model lungsPpeakPplatGapPattern
Baseline21.712.59.2
Resistance ×340.113.027.1Gap wide, plateau the same
Compliance ×0.528.219.09.2Both up, gap the same

Pressures in cmH₂O. Volume control, tidal volume 450 mL, rate 12/min, PEEP 5 cmH₂O, Ti 1.0 s with a 0.5 s pause.

Gap wide, plateau about the same: the airways

Something between the ventilator and the alveoli is narrower. Look for a kinked or bitten tube, secretions or a mucus plug, bronchospasm, and water or a blocked filter in the circuit. Suction, add a bite block, give a bronchodilator when wheeze points to bronchospasm, and check the circuit.

Plateau up as well: the lungs, the pleura, or trapped air

First do an expiratory hold. If total PEEP is well above the set PEEP, air is trapped (auto-PEEP). In model lungs with bronchospasm, raising the rate from 14 to 22/min raised Ppeak from 24.4 to 32.9 cmH₂O and total PEEP from 8.9 to 12.8 cmH₂O: the fix there is a slower rate, a longer expiration and treating the bronchospasm.

If total PEEP is not raised, the respiratory system is stiffer: mainstem intubation (check tube depth and listen to both sides), pneumothorax, pulmonary edema, worsening ARDS or pneumonia, atelectasis, or a tense abdomen. Examine the patient and use a chest x-ray or ultrasound.

Sudden or gradual?

Do not just raise the limit

Raising the alarm limit silences the alarm, not the cause, and lets higher pressure reach the lungs. In pressure control the pressure is set, so a stiffer lung shows up as a smaller tidal volume and a low volume alarm instead (pressure vs volume control).

Key point

Patient first. Then the plateau: a wide gap points to the tube and airways, a high plateau to trapped air or the lungs, pleura, chest wall and abdomen.

Try it

  1. Open the MV Simulator, choose Advanced, set Tpause to 0.5 s and press Confirm.
  2. Choose the Airway plug patient and compare Ppeak with Pplat. Then Stiff lungs. Which one moved the plateau?
  3. Choose Bronchospasm, raise the rate, and press Exp. hold to see the trapped pressure.

References

  1. Hess DR. Respiratory mechanics in mechanically ventilated patients. Respir Care 2014;59(11):1773-1794. doi:10.4187/respcare.03410
  2. Marini JJ. Dynamic hyperinflation and auto-positive end-expiratory pressure: lessons learned over 30 years. Am J Respir Crit Care Med 2011;184(7):756-762. doi:10.1164/rccm.201102-0226PP
  3. Leatherman JW, Ravenscraft SA. Low measured auto-positive end-expiratory pressure during mechanical ventilation of patients with severe asthma: hidden auto-positive end-expiratory pressure. Crit Care Med 1996;24(3):541-546. doi:10.1097/00003246-199603000-00028