Ventilator lessons Lesson 8 of 9
Adjusting ventilator settings from the ABG
Updated
On the ventilator, PaCO₂ is set by ventilation (rate and tidal volume) and PaO₂ by oxygenation (FiO₂ and PEEP). Read the blood gas, decide which problem you have, and change the setting that controls it. Then wait for a new steady state before judging the result.
Hands-on lesson · about 6 minARDS: PEEP up, then FiO₂ downChange one setting at a time and read the blood gas after each change.What to change
| Blood gas | First change | Watch |
|---|---|---|
| PaCO₂ high, pH low | Raise the rate. Raise tidal volume only if plateau pressure and mL/kg allow | Auto-PEEP, plateau pressure |
| PaCO₂ low, pH high | Lower the rate (or tidal volume) | A patient who triggers extra breaths keeps the rate up |
| PaO₂ or SpO₂ low | Raise FiO₂ and PEEP, in steps | Blood pressure, plateau and driving pressure |
| PaO₂ high | Lower FiO₂ first, then PEEP | SpO₂ after each step |
| Metabolic acidosis | Treat the cause. Keep minute ventilation at least as high as the patient was breathing before intubation; extra ventilation can only partly compensate | pH, the patient's own rate |
PaCO₂: alveolar ventilation
PaCO₂ follows alveolar ventilation, the part of each breath that reaches gas-exchanging lung: PaCO₂ = 0.863 × CO₂ production ÷ alveolar ventilation, and alveolar ventilation is rate × (tidal volume − dead space). If CO₂ production, tidal volume and dead space stay the same, PaCO₂ changes in inverse proportion to the rate. That gives a bedside estimate:
| New rate | Current rate × current PaCO₂ ÷ target PaCO₂ |
|---|
In this example, lowering tidal volume from 480 to 360 mL at rate 19/min raised PaCO₂ from 40.6 to 66.6 mmHg (pH 7.21) after an hour, and it was still creeping up. For a target near 50 mmHg the formula gives 19 × 66.6 ÷ 50 = 25.3/min. At 25/min the model settled at PaCO₂ 51.5 mmHg, pH 7.31. The formula assumes a steady state; a PaCO₂ that is still rising will underestimate the rate needed.
Minute ventilation was back near where it started (8.7 and 8.5 L/min), yet PaCO₂ stayed 11 mmHg higher. Each smaller breath spends a larger share on dead space, so the same minute ventilation clears less CO₂. That is why the formula is a reasonable estimate for rate changes but not for tidal volume changes.
PaCO₂ moves slowly. In this model 90% of the rise took about 43 minutes, so a blood gas drawn a few minutes after a change can mislead.
The limit on rate: trapped air
A faster rate shortens expiration. In model lungs with bronchospasm, raising the rate from 14 to 22/min lowered PaCO₂ from 53.8 to 33.8 mmHg but raised auto-PEEP from 3.9 to 7.8 cmH₂O. Going back to 14/min with a longer expiration (I:E 1:4) cut auto-PEEP to 2.8 cmH₂O at the price of PaCO₂ 57.7 mmHg, pH 7.27. In severe obstruction, a higher PaCO₂ is often accepted to protect against trapping (auto-PEEP).
PaO₂: FiO₂ and PEEP
| PEEP | FiO₂ | PaO₂ (mmHg) | Shunt |
|---|---|---|---|
| 5 | 0.5 | 67 | 37% |
| 10 | 0.5 | 79 | 29% |
| 15 | 0.5 | 107 | 19% |
| 15 | 0.4 | 89 | 20% |
| 15 | 0.3 | 76 | 19% |
| 5 | 1 | 117 | 35% |
| 15 | 1 | 388 | 19% |
Model ARDS lungs, tidal volume 450 mL, rate 20/min. Shunt is the model's fraction of blood passing unventilated lung.
With a large shunt, oxygen does not reach the blood flowing past collapsed alveoli, so FiO₂ helps little: at PEEP 5, pure oxygen gave PaO₂ 117 mmHg. PEEP that reopens lung lowers the shunt; at PEEP 15 the same oxygen gave 388 mmHg. Once PEEP has done its work, FiO₂ can come down. The hands-on PEEP lesson walks through these steps one at a time.
Targets in ARDS
- Oxygenation: PaO₂ 55 to 80 mmHg or SpO₂ 88 to 95% (the ARDS Network protocol).
- pH 7.30 to 7.45. Below 7.30, raise the rate until pH is above 7.30 or PaCO₂ is below 25 mmHg (maximum 35/min); a raised PaCO₂ is accepted to keep tidal volume and plateau pressure low.
- Plateau pressure 30 cmH₂O or less, tidal volume 6 mL/kg PBW (4 to 8).
PEEP and FiO₂ are often raised together using the ARDS Network tables (ARDSNet PEEP table).
Key point
Rate and tidal volume move PaCO₂; FiO₂ and PEEP move PaO₂. Change one, wait for the new steady state, then draw the next gas.
Try it
- Hands-on lesson: ARDS, PEEP up, then FiO₂ down
- Hands-on lesson: low tidal volume and the rate
- In the MV Simulator, change one setting and press Draw ABG after the fast-forward.
References
- NIH NHLBI ARDS Clinical Network. Mechanical ventilation protocol summary, 2008 (archived copy). web.archive.org
- Rawat D, Modi P, Sharma S. Hypercapnia. StatPearls. Treasure Island (FL): StatPearls Publishing; updated 2023. www.ncbi.nlm.nih.gov
- Acute Respiratory Distress Syndrome Network. Ventilation with lower tidal volumes as compared with traditional tidal volumes for acute lung injury and the acute respiratory distress syndrome. N Engl J Med 2000;342(18):1301-1308. doi:10.1056/NEJM200005043421801
- Fan E, Del Sorbo L, Goligher EC, et al. An official American Thoracic Society/European Society of Intensive Care Medicine/Society of Critical Care Medicine clinical practice guideline: mechanical ventilation in adult patients with acute respiratory distress syndrome. Am J Respir Crit Care Med 2017;195(9):1253-1263. doi:10.1164/rccm.201703-0548ST
- Tuxen DV, Lane S. The effects of ventilatory pattern on hyperinflation, airway pressures, and circulation in mechanical ventilation of patients with severe air-flow obstruction. Am Rev Respir Dis 1987;136(4):872-879. doi:10.1164/ajrccm/136.4.872