Ventilator lessons Ventilator quiz

Ventilator quiz: waveforms and settings

Updated

24 multiple-choice questions at three levels, from reading the three traces to choosing settings from a blood gas. Each answer explains every choice and links the lesson that covers it. Every figure and number comes from the TracingsMD ventilator model.

Pick an answer in your head, then open Show the answer.

Beginner (8)

1. These traces show one passive breath. Which mode is it?

Pressure, flow and volume of one passive breath: inspiratory flow is flat and pressure climbs until the end of inspiration.
  1. Pressure control
  2. Volume control
  3. Pressure support
  4. CPAP
Show the answerHide the answer

Answer B: Volume control. Inspiratory flow is flat (a square) and pressure keeps climbing until the end of inspiration: the ventilator is delivering a set volume at a set flow, and the pressure is the result.

  • A. Pressure control: Pressure control holds pressure flat and lets flow decay.
  • B. Volume control: Flat flow with a climbing pressure.
  • C. Pressure support: Pressure support needs the patient to trigger each breath; this patient is passive.
  • D. CPAP: CPAP gives no breaths of its own.

2. Same lungs, another mode. Which mode is it?

Pressure, flow and volume of one passive breath: pressure rises to a flat top and inspiratory flow starts high and decays.
  1. Volume control
  2. Volume control with a pause
  3. Airway pressure release
  4. Pressure control
Show the answerHide the answer

Answer D: Pressure control. Pressure rises to a flat top of about 15.0 cmH₂O and inspiratory flow starts high and decays as the lungs fill. The pressure is set; volume and flow are the result.

  • A. Volume control: Volume control has a square flow and a climbing pressure.
  • B. Volume control with a pause: A pause shows as zero flow at the end of inspiration, not a decaying flow.
  • C. Airway pressure release: Airway pressure release keeps a long high pressure with brief releases.
  • D. Pressure control: Flat pressure with decaying flow.

3. On the flow trace, what does flow below the zero line mean?

The same three traces; expiratory flow is drawn below the zero line.
  1. Gas leaving the lungs (expiration)
  2. A circuit leak
  3. The patient trying to breathe in
  4. The ventilator sucking gas out
Show the answerHide the answer

Answer A: Gas leaving the lungs (expiration). By convention flow into the patient is drawn above zero and flow out below it. Expiration is passive: flow jumps out, then decays toward zero as the lungs empty.

  • A. Gas leaving the lungs (expiration): Below zero is gas coming out.
  • B. A circuit leak: A leak shows as exhaled volume smaller than inspired volume, not as negative flow.
  • C. The patient trying to breathe in: An effort to breathe in would pull flow upward, toward inspiration.
  • D. The ventilator sucking gas out: In normal ventilation expiration is passive; the ventilator only opens the valve.

4. Which letter marks the plateau pressure?

Airway pressure of one breath with a pause; A marks the peak, B the flat part during the pause, C the pressure at the end of expiration.
  1. A
  2. B
  3. C
  4. None: a plateau needs an expiratory hold
Show the answerHide the answer

Answer B: B. B sits on the flat part after flow stops at the end of inspiration (12.5 cmH₂O here). A is the peak (21.7), which includes the pressure to push flow through the airways. C is PEEP at the end of expiration.

  • A. A: A is the peak, measured while gas is flowing.
  • B. B: B is the plateau, measured with no flow.
  • C. C: C is end-expiratory pressure (PEEP).
  • D. None: a plateau needs an expiratory hold: The plateau comes from an inspiratory pause; an expiratory hold measures total PEEP.

5. On volume control the lungs become stiffer. What do you expect to see?

  1. Tidal volume falls; pressure stays the same
  2. Peak rises; plateau stays the same
  3. Peak and plateau pressure rise; tidal volume stays the same
  4. No change
Show the answerHide the answer

Answer C: Peak and plateau pressure rise; tidal volume stays the same. Volume control keeps delivering the set volume, so the pressure has to rise. In the model, halving compliance raised the peak from 17.4 to 24.1 cmH₂O with tidal volume 432 and 432 mL.

  • A. Tidal volume falls; pressure stays the same: That is what pressure control does.
  • B. Peak rises; plateau stays the same: That is the resistance pattern.
  • C. Peak and plateau pressure rise; tidal volume stays the same: Set volume, so pressure rises.
  • D. No change: Something has to give: in volume control it is pressure.

6. The same change on pressure control. What do you expect?

  1. Tidal volume falls
  2. Pressure rises
  3. Both stay the same
  4. Rate rises
Show the answerHide the answer

Answer A: Tidal volume falls. Pressure control holds the set pressure, so less volume goes in. In the model tidal volume fell from 444 to 270 mL while pressure stayed at 15.0 cmH₂O. Watch the low tidal volume alarm.

  • A. Tidal volume falls: Less volume for the same pressure.
  • B. Pressure rises: Pressure is the set value in pressure control.
  • C. Both stay the same: Volume changes.
  • D. Rate rises: A passive patient does not change its rate.

7. Driving pressure is calculated as:

  1. Ppeak − PEEP
  2. Ppeak − Pplat
  3. Pplat ÷ PEEP
  4. Pplat − total PEEP
Show the answerHide the answer

Answer D: Pplat − total PEEP. Driving pressure is the pressure that stretches the respiratory system: plateau pressure minus total PEEP. It also equals tidal volume divided by static compliance.

  • A. Ppeak − PEEP: Ppeak includes the resistive pressure, so this overstates it.
  • B. Ppeak − Pplat: That gap reflects airway resistance.
  • C. Pplat ÷ PEEP: A ratio has no meaning here.
  • D. Pplat − total PEEP: Correct.

8. Which settings mainly change PaCO₂?

  1. FiO₂ and PEEP
  2. FiO₂ and rate
  3. Rate and tidal volume
  4. PEEP and inspiratory time
Show the answerHide the answer

Answer C: Rate and tidal volume. PaCO₂ follows alveolar ventilation, set by rate and tidal volume. FiO₂ and PEEP mainly change oxygenation.

  • A. FiO₂ and PEEP: These set oxygenation.
  • B. FiO₂ and rate: FiO₂ does not clear CO₂.
  • C. Rate and tidal volume: Ventilation sets PaCO₂.
  • D. PEEP and inspiratory time: These change oxygenation and timing, not CO₂ clearance directly.

Intermediate (8)

9. After a change, the pressure looks like this. What is the most likely cause?

Pressure of one breath before and after a change: the peak is much higher after, the plateau about the same.
  1. Pulmonary edema
  2. Pneumothorax
  3. Abdominal distension
  4. Secretions or a kinked tube
Show the answerHide the answer

Answer D: Secretions or a kinked tube. The peak rose from 21.7 to 40.1 cmH₂O while the plateau stayed near 13.0: the gap widened, so airway resistance rose. Look at the tube and airways.

  • A. Pulmonary edema: Edema stiffens the lungs and raises the plateau too.
  • B. Pneumothorax: A pneumothorax raises the plateau.
  • C. Abdominal distension: A tense abdomen raises the plateau.
  • D. Secretions or a kinked tube: A wider gap with the same plateau is resistance.

10. Here both peak and plateau rose by about the same amount. Which cause fits best?

Pressure of one breath before and after a change: peak and plateau are both higher after, by about the same amount.
  1. A kinked tube
  2. Pulmonary edema
  3. Secretions in the tube
  4. Bronchospasm without trapping
Show the answerHide the answer

Answer B: Pulmonary edema. The plateau rose from 12.5 to 19.0 cmH₂O and the gap stayed the same, so the respiratory system got stiffer. Edema, ARDS, pneumonia, pneumothorax, mainstem intubation or a tense abdomen do this.

  • A. A kinked tube: A kinked tube widens the gap.
  • B. Pulmonary edema: Stiffer lungs raise the plateau.
  • C. Secretions in the tube: Secretions widen the gap.
  • D. Bronchospasm without trapping: Bronchospasm widens the gap; it raises the plateau only through trapped air.

11. Ppeak 23.4, Pplat 17.8 cmH₂O, constant inspiratory flow 27 L/min. What is the airway resistance?

  1. 5.6 cmH₂O·s/L
  2. 0.21 cmH₂O·s/L
  3. 12.4 cmH₂O·s/L
  4. 52 cmH₂O·s/L
Show the answerHide the answer

Answer C: 12.4 cmH₂O·s/L. Resistance = (Ppeak − Pplat) ÷ flow in L/s. 27 L/min is 0.45 L/s, so (23.4 − 17.8) ÷ 0.45 = 12.4 cmH₂O·s/L.

  • A. 5.6 cmH₂O·s/L: That is the pressure gap, not yet divided by flow.
  • B. 0.21 cmH₂O·s/L: Flow was left in L/min; convert to L/s first.
  • C. 12.4 cmH₂O·s/L: Correct.
  • D. 52 cmH₂O·s/L: Peak alone includes the elastic pressure.

12. Exhaled tidal volume 431 mL, Ppeak 23.4, Pplat 17.8, total PEEP 5.3 cmH₂O. What is the static compliance?

  1. 34.5 mL/cmH₂O
  2. 23.8 mL/cmH₂O
  3. 24.2 mL/cmH₂O
  4. 81.3 mL/cmH₂O
Show the answerHide the answer

Answer A: 34.5 mL/cmH₂O. Static compliance = tidal volume ÷ (Pplat − total PEEP) = 431 ÷ 12.5 = 34.5 mL/cmH₂O, low as in ARDS.

  • A. 34.5 mL/cmH₂O: Correct.
  • B. 23.8 mL/cmH₂O: Using the peak gives dynamic compliance.
  • C. 24.2 mL/cmH₂O: PEEP must be subtracted.
  • D. 81.3 mL/cmH₂O: That divides by PEEP alone.

13. What does this flow trace show?

Flow over several breaths: expiratory flow is still well below zero when each new breath starts.
  1. Expiration is cut off before flow reaches zero: air trapping
  2. A circuit leak during expiration
  3. Normal passive expiration
  4. The patient triggering extra breaths
Show the answerHide the answer

Answer A: Expiration is cut off before flow reaches zero: air trapping. Expiratory flow is still about 8 L/min out of the lungs when the next breath starts, so the lungs never finish emptying. This is auto-PEEP. Measure it with an expiratory hold.

  • A. Expiration is cut off before flow reaches zero: air trapping: Flow does not return to zero.
  • B. A circuit leak during expiration: A leak shows as missing exhaled volume.
  • C. Normal passive expiration: Normal expiration returns to zero before the next breath.
  • D. The patient triggering extra breaths: A triggered breath shows a small pressure or flow dip before inspiration; there is none here.

14. During an expiratory hold the pressure rises from the set PEEP of 5 to 9.2 cmH₂O. What is the auto-PEEP?

  1. 9.2 cmH₂O
  2. 5.0 cmH₂O
  3. 4.2 cmH₂O
  4. 14.2 cmH₂O
Show the answerHide the answer

Answer C: 4.2 cmH₂O. Auto-PEEP is total PEEP minus set PEEP: 9.2 − 5 = 4.2 cmH₂O.

  • A. 9.2 cmH₂O: That is total PEEP.
  • B. 5.0 cmH₂O: That is the set PEEP.
  • C. 4.2 cmH₂O: Correct.
  • D. 14.2 cmH₂O: Adding them double counts.

15. The high pressure alarm sounds. The patient is stable, passive and well oxygenated. What next?

  1. Raise the alarm limit
  2. Add an inspiratory pause and compare Ppeak with Pplat
  3. Switch to pressure control to stop the alarm
  4. Sedate more and wait
Show the answerHide the answer

Answer B: Add an inspiratory pause and compare Ppeak with Pplat. Once the patient is safe, the plateau tells you where the pressure comes from: a wide gap points to the airways and tube, a high plateau to trapped air or stiffer lungs.

  • A. Raise the alarm limit: That silences the alarm, not the cause.
  • B. Add an inspiratory pause and compare Ppeak with Pplat: Find the cause.
  • C. Switch to pressure control to stop the alarm: Pressure control hides the problem as a falling tidal volume.
  • D. Sedate more and wait: Sedation may help with fighting the ventilator but does not find the cause.

16. On volume control, adding a 0.5 s pause inside the same 1.0 s inspiratory time raised the peak from 17.4 to 21.7 cmH₂O with no change in the lungs. Why?

  1. The pause makes the lungs stiffer
  2. The pause adds volume
  3. Auto-PEEP
  4. The same volume had to go in over half the time, so flow doubled
Show the answerHide the answer

Answer D: The same volume had to go in over half the time, so flow doubled. Inspiratory flow rose from 27 to 54 L/min. The resistive part of the peak rises with flow; the plateau does not.

  • A. The pause makes the lungs stiffer: Compliance did not change.
  • B. The pause adds volume: The set volume is the same.
  • C. Auto-PEEP: Expiration was unchanged.
  • D. The same volume had to go in over half the time, so flow doubled: Flow doubled.

Advanced (8)

17. ARDS lungs, rate 19/min. Tidal volume is lowered from 480 to 360 mL. What happens to PaCO₂ over the next hour?

  1. It falls, because dead space falls
  2. It stays the same, because the rate is the same
  3. It rises, because alveolar ventilation falls
  4. It rises only if FiO₂ is lowered
Show the answerHide the answer

Answer C: It rises, because alveolar ventilation falls. Smaller breaths at the same rate move less gas past dead space. In the model PaCO₂ rose from 40.6 to 66.6 mmHg and was still rising after an hour.

  • A. It falls, because dead space falls: Dead space stays about the same; its share of each breath grows.
  • B. It stays the same, because the rate is the same: Rate alone does not fix alveolar ventilation.
  • C. It rises, because alveolar ventilation falls: Correct.
  • D. It rises only if FiO₂ is lowered: FiO₂ does not change CO₂ clearance.

18. PaCO₂ is 66.6 mmHg at rate 19/min. You want about 50 mmHg without changing tidal volume. Which rate does the bedside formula suggest?

  1. 14/min
  2. 35/min
  3. 19/min
  4. 25/min
Show the answerHide the answer

Answer D: 25/min. New rate = 19 × 66.6 ÷ 50 = 25.3/min. In the model, rate 25/min gave PaCO₂ 51.5 mmHg.

  • A. 14/min: The ratio is upside down.
  • B. 35/min: No need to jump to the maximum.
  • C. 19/min: Unchanged rate leaves PaCO₂ where it is.
  • D. 25/min: Correct.

19. After the rate was raised, minute ventilation was close to the start (8.7 and 8.5 L/min), but PaCO₂ was 51.5, not 40.6 mmHg. Why?

  1. CO₂ production doubled
  2. Smaller breaths spend a larger share on dead space
  3. The blood gas was drawn too early
  4. PEEP was too high
Show the answerHide the answer

Answer B: Smaller breaths spend a larger share on dead space. Alveolar ventilation is rate × (tidal volume − dead space). With the same dead space, a 360 mL breath delivers a smaller fraction to gas-exchanging lung than a 480 mL breath.

  • A. CO₂ production doubled: Nothing in the model changed CO₂ production.
  • B. Smaller breaths spend a larger share on dead space: Correct.
  • C. The blood gas was drawn too early: The model had reached a steady state.
  • D. PEEP was too high: PEEP was unchanged.

20. Severe bronchospasm. Raising the rate from 14 to 22/min lowered PaCO₂ from 53.8 to 33.8 mmHg but raised auto-PEEP from 3.9 to 7.8 cmH₂O. What is the better next step?

  1. Slow the rate and lengthen expiration, accepting a higher PaCO₂
  2. Raise the rate further to bring PaCO₂ to normal
  3. Raise the tidal volume to clear more CO₂
  4. Lower the set PEEP to zero so the lungs empty
Show the answerHide the answer

Answer A: Slow the rate and lengthen expiration, accepting a higher PaCO₂. Trapped air raises lung volume and pressure and can lower blood pressure. In the model, rate 14/min with I:E 1:4 cut auto-PEEP to 2.8 cmH₂O, with PaCO₂ 57.7 mmHg and pH 7.27.

  • A. Slow the rate and lengthen expiration, accepting a higher PaCO₂: Correct.
  • B. Raise the rate further to bring PaCO₂ to normal: Less time to exhale means more trapping.
  • C. Raise the tidal volume to clear more CO₂: More volume to exhale means more trapping.
  • D. Lower the set PEEP to zero so the lungs empty: Set PEEP is not what traps air here; the short expiration is.

21. ARDS lungs. FiO₂ 1.0 at PEEP 5 gives PaO₂ 117 mmHg; FiO₂ 1.0 at PEEP 15 gives 388. What explains the difference?

  1. Hypoventilation at the lower PEEP
  2. A diffusion limit that more oxygen cannot overcome
  3. Higher cardiac output carrying more oxygen at higher PEEP
  4. Shunt: PEEP reopened collapsed lung so oxygen could reach that blood
Show the answerHide the answer

Answer D: Shunt: PEEP reopened collapsed lung so oxygen could reach that blood. Blood that passes collapsed alveoli never meets the oxygen, so raising FiO₂ helps little. The model's shunt fell from 35% to 19% with PEEP.

  • A. Hypoventilation at the lower PEEP: PaCO₂ was about the same (43.0 and 42.7 mmHg).
  • B. A diffusion limit that more oxygen cannot overcome: Diffusion limits respond well to more oxygen.
  • C. Higher cardiac output carrying more oxygen at higher PEEP: Higher PEEP tends to lower, not raise, cardiac output.
  • D. Shunt: PEEP reopened collapsed lung so oxygen could reach that blood: Correct.

22. PEEP went from 5 to 15 at the same tidal volume. Pplat rose from 17.8 to 27.2 and driving pressure fell from 12.5 to 11.7 cmH₂O. What does that tell you?

  1. Some lung opened: compliance rose a little
  2. The added PEEP overstretched the lungs
  3. PEEP had no effect on the lungs
  4. Driving pressure always falls when PEEP rises
Show the answerHide the answer

Answer A: Some lung opened: compliance rose a little. Driving pressure fell because compliance rose, from 34.5 to 36.9 mL/cmH₂O. If PEEP had only stretched open lung, compliance would fall and driving pressure would rise. In a patient, a change this small can be within measurement error.

  • A. Some lung opened: compliance rose a little: Correct.
  • B. The added PEEP overstretched the lungs: Overstretch would raise driving pressure.
  • C. PEEP had no effect on the lungs: Oxygenation and compliance both changed.
  • D. Driving pressure always falls when PEEP rises: It rises when PEEP overdistends.

23. Bronchospasm: Pplat 18.9, set PEEP 5, total PEEP on an expiratory hold 12.8 cmH₂O. What is the driving pressure?

  1. 13.9 cmH₂O
  2. 18.9 cmH₂O
  3. 6.1 cmH₂O
  4. 7.8 cmH₂O
Show the answerHide the answer

Answer C: 6.1 cmH₂O. Use total PEEP: 18.9 − 12.8 = 6.1 cmH₂O. Subtracting the set PEEP would overstate it at 13.9.

  • A. 13.9 cmH₂O: That uses the set PEEP and ignores the trapped pressure.
  • B. 18.9 cmH₂O: That is the plateau itself.
  • C. 6.1 cmH₂O: Correct.
  • D. 7.8 cmH₂O: That is the auto-PEEP.

24. In the ARDS Network protocol, what is the oxygenation goal?

  1. PaO₂ above 100 mmHg
  2. PaO₂ 55 to 80 mmHg or SpO₂ 88 to 95%
  3. SpO₂ 96 to 100%
  4. PaO₂ 80 to 120 mmHg
Show the answerHide the answer

Answer B: PaO₂ 55 to 80 mmHg or SpO₂ 88 to 95%. The protocol pairs FiO₂ and PEEP to keep PaO₂ 55 to 80 mmHg or SpO₂ 88 to 95%. Higher targets add oxygen exposure and pressure without a proven benefit.

  • A. PaO₂ above 100 mmHg: Higher than the goal.
  • B. PaO₂ 55 to 80 mmHg or SpO₂ 88 to 95%: Correct.
  • C. SpO₂ 96 to 100%: Higher than the goal.
  • D. PaO₂ 80 to 120 mmHg: Higher than the goal.

Keep practicing

Every trace on this page is computed by the TracingsMD ventilator model, not recorded from a patient. For learning only, not for diagnosis or patient care.

References

  1. NIH NHLBI ARDS Clinical Network. Mechanical ventilation protocol summary, 2008 (archived copy). web.archive.org
  2. Hess DR. Respiratory mechanics in mechanically ventilated patients. Respir Care 2014;59(11):1773-1794. doi:10.4187/respcare.03410
  3. Amato MBP, Meade MO, Slutsky AS, et al. Driving pressure and survival in the acute respiratory distress syndrome. N Engl J Med 2015;372(8):747-755. doi:10.1056/NEJMsa1410639
  4. Arnal JM, Garnero A, Saoli M, Chatburn RL. Parameters for simulation of adult subjects during mechanical ventilation. Respir Care 2018;63(2):158-168. doi:10.4187/respcare.05775