Ventilator lessons Lesson 1 of 9

Ventilator waveforms explained: pressure, flow and volume

Updated

A ventilator screen draws three traces on one time axis: airway pressure, flow and volume. Read them together. In each mode one of them is set by you and the others show how the lungs respond.

Three stacked traces over 10 seconds. Pressure rises from PEEP during inspiration. Flow is a flat square going in and a decaying curve coming out. Volume climbs and falls back to zero.
Two volume control breaths. Adult model patient, VC, set tidal volume 450 mL, rate 12/min, PEEP 5 cmH₂O, Ti 1.0 s, no pause.

Pressure: the pressure-time waveform

Airway pressure (Paw) starts each breath at PEEP, the pressure kept in the circuit at the end of expiration. Here it starts at 5 cmH₂O. In volume control it climbs throughout inspiration and reaches 17.4 cmH₂O, then falls back to PEEP once the exhalation valve opens.

Flow: the flow-time waveform

Flow above the zero line is gas going in; flow below it is gas coming out. In this volume control breath, inspiratory flow is a flat 27 L/min square, because the ventilator holds it constant. Expiration is passive: flow jumps to about 38 L/min, then decays toward zero as the lungs empty.

Expiratory flow is the trace to check first. If it does not reach zero before the next breath, the lungs have not finished emptying (auto-PEEP).

Volume: the volume-time waveform

Volume rises during inspiration and falls during expiration. It is drawn here as the amount above end-expiration, so each breath should return to the same baseline. In this recording the measured tidal volume is 432 mL, slightly below the set 450 mL.

A reading order

  1. Read the mode on the screen, then match the shapes. Flat flow with climbing pressure is volume control. Flat pressure with decaying flow is usually pressure control, although volume-targeted pressure modes look the same (pressure vs volume control).
  2. Follow expiratory flow to the end. Does it reach zero before the next breath?
  3. Check that volume returns to baseline. If it does not, suspect a leak or trapping that is still building. A steady baseline does not rule out trapping.
  4. Ask what the pressure means. A peak alone does not separate airway resistance from stiff lungs; a plateau does (peak vs plateau).

Key point

The trace you did not set is the one that tells you about the patient.

Words on the screen

Try it

  1. Open the MV Simulator. It starts in volume control.
  2. Watch one breath on the three traces: pressure climbs, flow is flat going in and decays coming out, and volume rises and returns to baseline.
  3. Select the VT tile, set a smaller volume and press Confirm. Over the next breaths the pressure top and the volume trace both fall.

Test yourself: ventilator quiz.

What these traces do not show

The model patient is passive and does not breathe on its own. Real traces can also show patient effort, asynchrony, circuit leaks and secretions. The model does not produce those patterns, so these lessons do not draw them.

References

  1. Hess DR. Respiratory mechanics in mechanically ventilated patients. Respir Care 2014;59(11):1773-1794. doi:10.4187/respcare.03410
0 of 9 lessons complete

Saving is off. Marks last for this page only.

Every trace on this page is computed by the TracingsMD ventilator model, not recorded from a patient. For learning only, not for diagnosis or patient care.