Ventilator lessons Lesson 2 of 9
Peak vs plateau pressure on the ventilator
Updated
Peak pressure is measured while gas is flowing; plateau pressure is measured with no flow. At a constant inspiratory flow, the gap between them follows airway resistance. Plateau pressure above PEEP follows the stiffness of the respiratory system.
What each number is
- Ppeak is the highest pressure in the breath. It includes the pressure needed to push flow through the airways.
- Pplat is read at the end of a pause with no flow, so the resistive part is gone. What is left is the pressure holding the delivered volume in the lungs and chest wall, plus PEEP.
- Driving pressure is Pplat minus total PEEP (driving pressure).
The arithmetic on this breath
| Resistance | (Ppeak − Pplat) ÷ flow = 9.2 cmH₂O ÷ 0.90 L/s = 10.2 cmH₂O·s/L |
|---|---|
| Compliance | Tidal volume ÷ driving pressure = 414 mL ÷ 7.5 cmH₂O = 55 mL/cmH₂O |
Two ways to get a high peak
| Ppeak | Pplat | Peak-plateau gap | Resistance | Compliance | |
|---|---|---|---|---|---|
| Baseline | 21.7 | 12.5 | 9.2 | 10.2 | 55 |
| Resistance ×3 | 40.1 | 13.0 | 27.0 | 30.0 | 56 |
| Compliance ×0.5 | 28.2 | 19.0 | 9.1 | 10.1 | 30 |
Pressures in cmH₂O, resistance in cmH₂O·s/L, compliance in mL/cmH₂O.
| Pattern | Points toward | Common causes |
|---|---|---|
| Peak up, plateau about the same | Higher resistance | Kinked or bitten tube, secretions, bronchospasm, narrow tube |
| Peak and plateau both up | Lower compliance or air trapping | Pulmonary edema, ARDS, pneumonia, atelectasis, pneumothorax, mainstem intubation, abdominal distension, stiff chest wall, auto-PEEP |
Getting a reliable plateau
- Use a pause of about 0.5 s or an inspiratory hold, in a patient who is not breathing against the ventilator. Patient effort makes the reading unreliable.
- Flow affects the peak but not the plateau. In this model a 0.5 s pause inside the same 1.0 s Ti doubles the inspiratory flow, from 27 to 54 L/min, and the peak rises from 17.4 to 21.7 cmH₂O with no change in the lungs. Measured tidal volume is 414 mL with the pause and 432 mL without it.
- Pplat reflects the lungs and the chest wall together. A stiff chest wall or a full abdomen can raise it without any change in the lungs.
Why the plateau number matters
The ARDS Network trial limited plateau pressure to 30 cmH₂O or less alongside a tidal volume of 6 mL/kg predicted body weight, and mortality was lower than with traditional tidal volumes. In a later analysis of ARDS trials, lower driving pressure was associated with better survival. That analysis shows an association; it did not test a driving pressure target. The 2023 ESICM guideline continues to recommend low tidal volumes in ARDS.
Key point
When the peak alarm sounds, look at the patient first. If the patient is unstable, disconnect and ventilate by hand. If the patient is stable and passive, get a plateau. A wide gap sends you to the tube and airways. A high plateau sends you to the lungs, pleura, chest wall and abdomen; in an obstructed patient it can be trapped air, so add an expiratory hold.
Step by step: high pressure alarm. The formulas: compliance and resistance.
Try it
- Open the MV Simulator, select the Tpause tile, set 0.5 s and press Confirm.
- Under Patient events, in the Lungs row, set Resistance to ×3 and press Apply. Watch the gap open while Pplat barely moves.
- Press Back to its own lungs. Then set Compliance to ×0.50 and press Apply. Watch Ppeak and Pplat rise together.
References
- Hess DR. Respiratory mechanics in mechanically ventilated patients. Respir Care 2014;59(11):1773-1794. doi:10.4187/respcare.03410
- Acute Respiratory Distress Syndrome Network. Ventilation with lower tidal volumes as compared with traditional tidal volumes for acute lung injury and the acute respiratory distress syndrome. N Engl J Med 2000;342(18):1301-1308. doi:10.1056/NEJM200005043421801
- Amato MBP, Meade MO, Slutsky AS, et al. Driving pressure and survival in the acute respiratory distress syndrome. N Engl J Med 2015;372(8):747-755. doi:10.1056/NEJMsa1410639
- Grasselli G, Calfee CS, Camporota L, et al. ESICM guidelines on acute respiratory distress syndrome: definition, phenotyping and respiratory support strategies. Intensive Care Med 2023;49(7):727-759. doi:10.1007/s00134-023-07050-7
Saving is off. Marks last for this page only.
Save progress on this device?
Every trace on this page is computed by the TracingsMD ventilator model, not recorded from a patient. For learning only, not for diagnosis or patient care.