Ventilator lessons Lesson 2 of 9

Peak vs plateau pressure on the ventilator

Updated

Peak pressure is measured while gas is flowing; plateau pressure is measured with no flow. At a constant inspiratory flow, the gap between them follows airway resistance. Plateau pressure above PEEP follows the stiffness of the respiratory system.

Airway pressure rises to a peak, drops to a flat plateau during a no-flow pause, then falls to PEEP. Flow below shows a square going in, then zero during the pause.
One breath with a 0.5 s inspiratory pause. Adult model patient, VC, set tidal volume 450 mL, rate 12/min, PEEP 5 cmH₂O, Ti 1.0 s. Values are the model's own measurements.

What each number is

The arithmetic on this breath

Resistance(Ppeak − Pplat) ÷ flow = 9.2 cmH₂O ÷ 0.90 L/s = 10.2 cmH₂O·s/L
ComplianceTidal volume ÷ driving pressure = 414 mL ÷ 7.5 cmH₂O = 55 mL/cmH₂O

Two ways to get a high peak

Three pressure traces side by side. With resistance tripled the peak is far higher but the plateau is nearly unchanged. With compliance halved both peak and plateau rise.
The same breath after one change to the model's lungs. Tidal volume and flow are the same in all three.
PpeakPplatPeak-plateau gapResistanceCompliance
Baseline21.712.59.210.255
Resistance ×340.113.027.030.056
Compliance ×0.528.219.09.110.130

Pressures in cmH₂O, resistance in cmH₂O·s/L, compliance in mL/cmH₂O.

PatternPoints towardCommon causes
Peak up, plateau about the sameHigher resistanceKinked or bitten tube, secretions, bronchospasm, narrow tube
Peak and plateau both upLower compliance or air trappingPulmonary edema, ARDS, pneumonia, atelectasis, pneumothorax, mainstem intubation, abdominal distension, stiff chest wall, auto-PEEP

Getting a reliable plateau

Why the plateau number matters

The ARDS Network trial limited plateau pressure to 30 cmH₂O or less alongside a tidal volume of 6 mL/kg predicted body weight, and mortality was lower than with traditional tidal volumes. In a later analysis of ARDS trials, lower driving pressure was associated with better survival. That analysis shows an association; it did not test a driving pressure target. The 2023 ESICM guideline continues to recommend low tidal volumes in ARDS.

Key point

When the peak alarm sounds, look at the patient first. If the patient is unstable, disconnect and ventilate by hand. If the patient is stable and passive, get a plateau. A wide gap sends you to the tube and airways. A high plateau sends you to the lungs, pleura, chest wall and abdomen; in an obstructed patient it can be trapped air, so add an expiratory hold.

Step by step: high pressure alarm. The formulas: compliance and resistance.

Try it

  1. Open the MV Simulator, select the Tpause tile, set 0.5 s and press Confirm.
  2. Under Patient events, in the Lungs row, set Resistance to ×3 and press Apply. Watch the gap open while Pplat barely moves.
  3. Press Back to its own lungs. Then set Compliance to ×0.50 and press Apply. Watch Ppeak and Pplat rise together.

References

  1. Hess DR. Respiratory mechanics in mechanically ventilated patients. Respir Care 2014;59(11):1773-1794. doi:10.4187/respcare.03410
  2. Acute Respiratory Distress Syndrome Network. Ventilation with lower tidal volumes as compared with traditional tidal volumes for acute lung injury and the acute respiratory distress syndrome. N Engl J Med 2000;342(18):1301-1308. doi:10.1056/NEJM200005043421801
  3. Amato MBP, Meade MO, Slutsky AS, et al. Driving pressure and survival in the acute respiratory distress syndrome. N Engl J Med 2015;372(8):747-755. doi:10.1056/NEJMsa1410639
  4. Grasselli G, Calfee CS, Camporota L, et al. ESICM guidelines on acute respiratory distress syndrome: definition, phenotyping and respiratory support strategies. Intensive Care Med 2023;49(7):727-759. doi:10.1007/s00134-023-07050-7