Ventilator lessons Lesson 6 of 9

Compliance and resistance on the ventilator: how to calculate them

Updated

Two numbers describe a passive breath: compliance, the volume you get for each cmH₂O of stretch, and resistance, the pressure it takes to push flow through the airways. Both come from peak pressure, plateau pressure, PEEP, tidal volume and inspiratory flow.

MV SimulatorMeasure them on the modelAdd a pause, read Ppeak and Pplat, then switch to bronchospasm and calculate again.

The formulas

ValueFormulaUnit
Static compliance (Cstat)Tidal volume ÷ (Pplat − total PEEP)mL/cmH₂O
Dynamic compliance (Cdyn)Tidal volume ÷ (Ppeak − total PEEP)mL/cmH₂O
Airway resistance (R)(Ppeak − Pplat) ÷ inspiratory flowcmH₂O·s/L
Time constant (τ)R × Cstat (with compliance in L/cmH₂O)s

Resistance needs a constant inspiratory flow, as in volume control with a square flow. Flow in L/min ÷ 60 gives L/s.

Worked example: model ARDS lungs

Pressure of one breath with peak, plateau and total PEEP marked, and below it a flat inspiratory flow of 27 liters per minute.
Volume control, tidal volume 450 mL, rate 20/min, PEEP 5 cmH₂O, Ti 1.5 s with a 0.5 s pause. Ppeak is read from the trace; Pplat and total PEEP are the model's hold measurements.
MeasuredPpeak 23.4, Pplat 17.8, total PEEP 5.3 cmH₂O, exhaled tidal volume 431 mL, flow 0.45 L/s
Static compliance431 ÷ (17.8 − 5.3) = 34.5 mL/cmH₂O
Dynamic compliance431 ÷ (23.4 − 5.3) = 23.8 mL/cmH₂O
Resistance(23.4 − 17.8) ÷ 0.45 = 12.4 cmH₂O·s/L
Time constant12.4 × 0.0345 = 0.43 s

The Ppeak in the table is from the hold breath. The trace above has a slightly higher peak and a 0.5 s pause plateau of 18.2 cmH₂O, because pressure keeps settling after flow stops. Use a pause long enough for the plateau to go flat.

Compare: model lungs with bronchospasm

ARDSBronchospasm
Ppeak / Pplat23.4 / 17.824.4 / 15.5
Total PEEP5.38.9 (set 5)
Inspiratory flow0.45 L/s0.32 L/s
Static compliance34.566.5
Resistance12.427.8
Time constant0.43 s1.85 s

Pressures in cmH₂O, compliance in mL/cmH₂O, resistance in cmH₂O·s/L. Bronchospasm: volume control 450 mL, rate 14/min, I:E 1:2, PEEP 5.

The ARDS lungs are stiff with near-normal airways: low compliance, short time constant, so they empty quickly. The bronchospasm lungs are compliant with narrow airways: high resistance and a time constant 4.3 times longer. Exhaling takes about three time constants, so those lungs need a long expiration, or they trap air. Expiratory resistance is usually higher than inspiratory resistance, so the real expiratory time constant can be longer than R × C suggests (auto-PEEP).

Typical values

In a study of 359 deeply sedated, passively ventilated adults, median static compliance was 54 mL/cmH₂O with normal lungs, 39 in ARDS and 59 in COPD. Median inspiratory resistance was 13, 12 and 22 cmH₂O·s/L, and the expiratory time constant 0.60, 0.46 and 1.07 s. The model ARDS lungs above sit close to those ARDS values.

Static or dynamic compliance?

Dynamic compliance uses the peak, so it falls when resistance rises even if the lungs are unchanged. Static compliance uses the plateau and reflects only the stiffness of the lungs and chest wall. When dynamic compliance falls, check whether static compliance fell too: if not, the problem is in the airways or the tube.

What lowers compliance and what raises resistance

ChangeCommon causes
Lower complianceARDS, pulmonary edema, pneumonia, atelectasis, fibrosis, pneumothorax, mainstem intubation, overdistension at high volume or PEEP, stiff chest wall, abdominal distension, auto-PEEP
Higher resistanceNarrow, kinked or bitten tube, secretions, bronchospasm, asthma or COPD, a wet or blocked filter

Getting numbers you can trust

Key point

The peak-plateau gap is the airways; the plateau above PEEP is the lungs and chest wall. Compliance times resistance tells you how long the lungs need to empty.

Try it

  1. Open the MV Simulator, choose Advanced, select the Tpause tile, set 0.5 s and press Confirm.
  2. Read Ppeak, Pplat, PEEP and VTe, and calculate compliance and resistance as above.
  3. Choose the Bronchospasm patient and calculate again. Watch the gap widen.

Related: peak vs plateau pressure, driving pressure.

References

  1. Arnal JM, Garnero A, Saoli M, Chatburn RL. Parameters for simulation of adult subjects during mechanical ventilation. Respir Care 2018;63(2):158-168. doi:10.4187/respcare.05775
  2. Hess DR. Respiratory mechanics in mechanically ventilated patients. Respir Care 2014;59(11):1773-1794. doi:10.4187/respcare.03410
  3. Marini JJ. Dynamic hyperinflation and auto-positive end-expiratory pressure: lessons learned over 30 years. Am J Respir Crit Care Med 2011;184(7):756-762. doi:10.1164/rccm.201102-0226PP
0 of 9 lessons complete

Saving is off. Marks last for this page only.

Every trace on this page is computed by the TracingsMD ventilator model, not recorded from a patient. For learning only, not for diagnosis or patient care.