Ventilator lessons Lesson 4 of 9
Pressure control vs volume control waveforms
Updated
In volume control you set the volume and the pressure is the result. In pressure control you set the pressure and the volume is the result. The waveforms show which is which at a glance.
| Volume control | Pressure control | |
|---|---|---|
| You set | Tidal volume, flow or inspiratory time | Inspiratory pressure, inspiratory time |
| Pressure trace | Climbs to a peak | Rises to a flat top |
| Inspiratory flow | Constant, a square (some ventilators also offer a ramp) | Starts high and decays |
| Varies with the lungs | Pressure | Tidal volume and flow |
| Watch for | High peak and plateau pressure | Tidal volume too low or too high |
The same lungs in each mode
Tidal volume is close in the two recordings (432 and 444 mL), but the shapes differ. Volume control reaches a peak of 17.4 cmH₂O with a flat 27 L/min. Pressure control holds 15.0 cmH₂O, and its flow starts at 48 L/min and falls as the lungs fill.
What happens when the lungs get stiffer
| Compliance ×0.5 | Peak pressure | Tidal volume |
|---|---|---|
| Volume control | 17.4 → 24.1 cmH₂O | 432 → 432 mL |
| Pressure control | 15.0 → 15.0 cmH₂O | 444 → 270 mL |
The change shows up in a different place in each mode. In volume control the pressure alarm is the one that speaks. In pressure control, pressure looks reassuringly flat while tidal volume falls, so set the low and high volume alarms with care. Volume can also rise, with patient effort or as the lungs improve.
Which is better?
Neither mode has been shown to be better overall. A systematic review of the two modes found no clear difference in outcomes, although the trials were small and varied in quality. The choice usually comes down to what you want to keep fixed. Whichever you use, check both the tidal volume and the plateau pressure.
Key point
Whatever you set stays still on the screen. Watch the trace that is free to move.
Try it
- Open the MV Simulator. It starts in volume control. Note Ppeak and VTe.
- Press the Mode button, choose PC-CMV, set Inspiratory pressure to 10 and press Confirm mode. Small differences from this page are normal.
- Under Patient events, in the Lungs row, set Compliance to ×0.50 and press Apply. Watch VTe fall while pressure holds. The low volume alarms come on once live breaths return (after the fast-forward, if it is on). Low minute volume stays on; Low tidal volume can come and go while VTe sits near its limit. In PC the limits here are set from the breaths before the event, so they do not follow the patient down.
For a guided version, run the ARDS in pressure control case.
References
- Rittayamai N, Katsios CM, Beloncle F, et al. Pressure-controlled vs volume-controlled ventilation in acute respiratory failure: a physiology-based narrative and systematic review. Chest 2015;148(2):340-355. doi:10.1378/chest.14-3169
- Hess DR. Respiratory mechanics in mechanically ventilated patients. Respir Care 2014;59(11):1773-1794. doi:10.4187/respcare.03410