Auto-PEEP on ventilator waveforms

Updated

Hands-on lesson · about 4 min

Air trapping (auto-PEEP)

Severe airflow obstruction on volume control. Raise the rate, then slow it and lengthen expiration, and watch the flow trace and intrinsic PEEP. Every number comes from a recorded run of the physiology model.

Changes appear faster here than in patients. After each change the lesson fast-forwards so you see the result within a minute. In patients oxygenation settles over minutes and CO₂ over 10 to 30 minutes.

Auto-PEEP means the next breath starts before the lungs have finished emptying. Pressure stays trapped in the lungs above the PEEP you set. It is also called intrinsic PEEP, and the trapped volume is called dynamic hyperinflation.

Spot it on the flow trace

Follow expiratory flow to its end. If it is still below zero when the next breath begins, the lungs were still emptying.

Two flow traces during bronchospasm. At rate 20 per minute the expiratory flow is well below zero when the next breath starts. At rate 10 per minute it has almost reached zero.
The same model lungs with bronchospasm (Severe in the Simulator) in volume control, set tidal volume 450 mL, Ti 1.0 s, PEEP 5 cmH₂O, at two rates.

At 20 breaths per minute, each 3 s cycle leaves 2 s to breathe out, and flow is still 7.6 L/min when the next breath starts. At 10 per minute there are 5 s, and it has fallen to 0.9 L/min.

The flow trace shows that trapping is present. It does not show how much, and the volume trace can hide it: at rate 20, every breath returns to the same baseline on the trace, yet the lungs end each breath about 220 mL fuller than the same lungs without bronchospasm.

Auto-PEEP calculation: measure it with an expiratory hold

An end-expiratory hold closes the circuit just as the next breath is due. With no flow, pressure in the airway and lungs equalizes, and the screen reads total PEEP. The model equalizes almost at once. In patients, read the value after the pressure stops rising, which can take a few seconds.

Airway pressure during an expiratory hold. Pressure starts at the set PEEP of 5 and rises to a higher flat level while flow stays at zero, then the next breath is released.
A 3 s expiratory hold at rate 20/min, same model lungs. Values are the model's own measurements.
Total PEEP9.2 cmH₂O
Set PEEP5.0 cmH₂O
Auto-PEEP (total minus set)4.2 cmH₂O

Just before the hold the airway pressure read 5.0 cmH₂O. The PEEP value on the screen shows the set pressure, not what is trapped.

The Simulator also shows lower estimates at these settings: iPEEP 2.6 cmH₂O under More values before a hold, and "Model estimate before the hold: 7.6 cmH₂O" under the result in the hold panel. Those are model estimates; the hold result is the measurement.

A hold needs a patient who is not breathing against the ventilator. In real patients it can read low when some airways close before the end of expiration, as in severe asthma. So a low reading, or flow that reaches zero, does not rule out trapping, and a high plateau pressure in that setting should raise suspicion.

Why it matters

What makes it better

When this triggers the high pressure alarm, see what to check. There is a cost: minute ventilation fell from 8.5 to 4.3 L/min between these two rates, so CO₂ clearance falls as well (adjusting the ventilator from the ABG).

Key point

Expiratory flow that never reaches zero is the warning sign. The expiratory hold gives the number.

Try it

  1. Open the MV Simulator, select the RR tile, set 20 and press Confirm.
  2. Under Patient events, set Bronchospasm to Severe and press Apply.
  3. Wait about 30 seconds, then press Exp. hold.
  4. Read Total PEEP (hold) in the Expiratory hold panel below the monitor, not the model estimate under it. Small differences from this page are normal.

For a guided version, run the airflow obstruction case: watch the first breaths as its brief asks, then change the rate and I:E. The case shows the model estimate and has no hold button, so measure total PEEP in the Simulator.

References

  1. Marini JJ. Dynamic hyperinflation and auto-positive end-expiratory pressure: lessons learned over 30 years. Am J Respir Crit Care Med 2011;184(7):756-762. doi:10.1164/rccm.201102-0226PP
  2. Tuxen DV, Lane S. The effects of ventilatory pattern on hyperinflation, airway pressures, and circulation in mechanical ventilation of patients with severe air-flow obstruction. Am Rev Respir Dis 1987;136(4):872-879. doi:10.1164/ajrccm/136.4.872
  3. Leatherman JW, Ravenscraft SA. Low measured auto-positive end-expiratory pressure during mechanical ventilation of patients with severe asthma: hidden auto-positive end-expiratory pressure. Crit Care Med 1996;24(3):541-546. doi:10.1097/00003246-199603000-00028
  4. Hess DR. Respiratory mechanics in mechanically ventilated patients. Respir Care 2014;59(11):1773-1794. doi:10.4187/respcare.03410
0 of 9 lessons complete

Saving is off. Marks last for this page only.

Every trace on this page is computed by the TracingsMD ventilator model, not recorded from a patient. For learning only, not for diagnosis or patient care.