Ventilator simulator

Ventilation & blood gases
Controls
PatientARDS: see the PEEP lesson

Monitor model values

HR--/min
SpO₂--%
MAP--mmHg
EtCO₂--mmHg

AdultPBW --
Breath--
Connecting…-Waiting for the engine
HR--
SpO₂ %--
MAP--
EtCO₂--

cmH₂OFixed
L/min · In + / Ex −Fixed
mLFixed
-- mL/kg PBWAirway pressure -- cmH₂O
P-V loop
F-V loop

Settings

Settings of the active mode
Waiting for settings

Inspiratory pauseNo qualified pause

Waiting for engine observations.

Waiting for completed breaths

Waiting for timingTiming details in Model information
Model view · absolute lung volume

A model value, not a ventilator reading: the ventilator screen above shows tidal volume.

Lung volumeL · absoluteFixedNow - L
Educational simulation · not for diagnosis.

Connecting to the physiological engine…

References

7 sources

The mechanics, holds and teaching notes on this page follow these sources.

  1. Hess DR. Respiratory mechanics in mechanically ventilated patients. Respir Care 2014;59(11):1773-1794. doi:10.4187/respcare.03410
  2. Marini JJ. Dynamic hyperinflation and auto-positive end-expiratory pressure: lessons learned over 30 years. Am J Respir Crit Care Med 2011;184(7):756-762. doi:10.1164/rccm.201102-0226PP
  3. Leatherman JW, Ravenscraft SA. Low measured auto-positive end-expiratory pressure during mechanical ventilation of patients with severe asthma: hidden auto-positive end-expiratory pressure. Crit Care Med 1996;24(3):541-546. doi:10.1097/00003246-199603000-00028
  4. Tuxen DV, Lane S. The effects of ventilatory pattern on hyperinflation, airway pressures, and circulation in mechanical ventilation of patients with severe air-flow obstruction. Am Rev Respir Dis 1987;136(4):872-879. doi:10.1164/ajrccm/136.4.872
  5. Rittayamai N, Katsios CM, Beloncle F, et al. Pressure-controlled vs volume-controlled ventilation in acute respiratory failure: a physiology-based narrative and systematic review. Chest 2015;148(2):340-355. doi:10.1378/chest.14-3169
  6. Acute Respiratory Distress Syndrome Network. Ventilation with lower tidal volumes as compared with traditional tidal volumes for acute lung injury and the acute respiratory distress syndrome. N Engl J Med 2000;342(18):1301-1308. doi:10.1056/NEJM200005043421801
  7. Amato MBP, Meade MO, Slutsky AS, et al. Driving pressure and survival in the acute respiratory distress syndrome. N Engl J Med 2015;372(8):747-755. doi:10.1056/NEJMsa1410639
START HERE · EXPLANATION

What the exercise shows

The settings bar holds your settings. VTe, minute volume and every waveform are model outputs. In Volume Control the ventilator delivers a set volume, so airway pressure is whatever the respiratory system needs to take that volume: a smaller VT needs less pressure, and the top of the Paw trace falls. The delivered VT can differ a little from the set value.

Teaching exercises on a model, not clinical guidance.

Waveform scale

Fixed scales are the default, so a change in pressure, flow or volume stays visible. Auto fits all visible samples with headroom: it expands immediately and waits at least 5 seconds before shrinking. Flow is positive during inspiration and negative during expiration. The loops have a frame of their own: it starts at the fixed scale, grows at once when a loop would leave it and shrinks back after 5 seconds. The breath in progress is drawn as it happens, with a dot at the newest sample, over the previous three breaths in fading lines.

Display only · patient state continues.

Ventilation mode

Choose a mode and review its settings. Nothing changes until you confirm the mode.

Waiting for the engine's supported settings.

The running patient state is preserved.
MV SIMULATOR

Alarm limits

Each bar shows a limit range with the current measured value as an arrow. Change a limit, or clear it to follow the settings again. Alarms are computed by this page from the measured values; the model does not raise them.

Alarm log model time

  1. No alarms yet.
Teaching alarms on a model, not a monitoring device.
MEASURED VALUE

Value

MV SIMULATOR

Patient

Patient in this session

The patient appears once the session opens.

Start with another adult

Enter sex, age, height and weight. The simulator uses the closest match among its model patients prepared in advance and shows the values it actually uses, with a line for each difference. The model calculates ideal body weight from sex and height, and the session starts on that patient's own settings.

Changing the patient starts a new session. The current traces and values are cleared.

What the patient model does not include

Model information

No engine connected.

Capabilities are checked before the session opens.

About this simulator

Physiology powered by Pulse Physiology Engine 4.3.2, distributed under the Apache License 2.0. These model observations are educational and are not clinical measurements.

Waveforms and blood values all come from one continuous simulated patient. What you set, what the model delivers and the blood values are shown as separate quantities.

The ventilator screen

The dark screen is a generic ventilator layout: mode and alarms on top, sweep waveforms on the left, measured values, Lung view and loops on the right, and settings at the bottom. Waveforms are drawn from the engine's original 50 Hz samples, left to right, with a small erased gap ahead of the newest sample; fixed scales are the default so changes stay visible. The volume waveform is tidal volume: lung volume above the end-expiratory lung volume of each breath.

Settings change like on a ventilator: select a tile, adjust it (drag, mouse wheel or arrow keys, Shift for bigger steps, or the slider on a phone), then Confirm or Cancel. An unconfirmed change goes back after 60 seconds. A new mode is reviewed in its own window and applied with Confirm mode. Notes about tidal volume per kg, PEEP plus inspiratory pressure, inverse I:E and short expiration never block a change; only the engine's own limits do.

What the numbers mean

Minute volume (MVe), respiratory rate, intrinsic PEEP (estimate) and end-tidal CO₂ are model ventilator outputs. Ppeak, measured PEEP, VTi, mean airway pressure over 6 breaths and measured I:E are read from the original samples of complete breaths, and so is VTe on the Simulator (in Cases, VTe is the model ventilator value). Pplat, driving pressure (DP), Cstat and Raw need a VC inspiratory pause and show -- without one. VT is the model's delivered tidal volume, separate from the set VT and from absolute lung volume (all the gas in the lungs), which is shown only in Model view. PBW is predicted body weight; for the model patient it equals the ideal body weight. Select any number for its definition and formula.

Arterial values are continuous model values, not collected blood samples. The bicarbonate basis is identified once the engine connects.

Alarms, Lung view and blood gases

Alarms are computed by this page from the measured values and the limits in Alarm limits; the model itself raises no alarms. High priority flashes red, medium flashes amber and low stays cyan, in the status bar only (no flashing when your system asks for reduced motion). Sound is off unless you turn on Alarm sound. SpO₂, heart rate and mean arterial pressure come from the model with every frame and are averaged like a bedside monitor (4, 8 and 2 seconds).

Lung view draws one large alveolus that inflates in place. Size: the air in the lung above residual volume. Residual volume: the air you cannot breathe out. Solid ring: the end of the last breath out; PEEP (plus any trapped air) sets it, and it stays while the next breath goes in. Dotted ring: one breath of 6 mL/kg PBW on top of the solid ring, so it moves with PEEP. Faint dashed ring: the model patient's healthy lung at rest without PEEP (FRC before any added lung condition), a reference and not a goal; a stiff or injured lung can end each breath out below it. The drawing stops at the patient's total lung capacity. Pink dashed ring while you adjust a VT or PEEP tile: a preview before Confirm. For VT, the end of the breath out plus the new VT (end of the breath in). For PEEP, the end of the breath out plus Cstat times the PEEP change, shown only after Cstat has been measured. It is an estimate that assumes compliance stays as last measured: in a real lung, recruitment can make the change larger and overdistension smaller. After Confirm it stays until one whole breath at the new setting has finished; then the solid ring (PEEP) or the alveolus at the end of each breath in (VT) shows the model's result. Color: the last plateau pressure, blue below 25 cmH₂O, turning amber from 25 to 30 and red from 30; gray until a pause (Tpause or an inspiratory hold) measures one, and again after a setting change. Pplat includes the chest wall. A driving pressure of 15 cmH₂O or more turns the DP label and the alveolus wall red. The color is a pressure cue only. One lung compartment: the view shows how much air the lung holds, not regional collapse or overstretch. The amber outline above 8 mL/kg PBW is a display threshold only: the model does not simulate overdistension injury. The arc shows airway pressure from 0 to 40 cmH₂O; the span from total PEEP to the last Pplat is the driving pressure. A patient without lung volumes shows the older picture instead: the alveolus fills with each breath relative to the reference tidal volume, and the dashed ring is 100 %.

On the Simulator page, Draw ABG copies the model's arterial values at that moment. As with a bedside blood gas analyzer, the result appears after 30 seconds of simulation time, in the Blood gas panel; a sample still in the analyzer when the session ends is not analyzed. The panel's Live tab shows the continuous model values, which at the bedside you only see through a drawn sample. P/F ratio, A-a gradient (PAO₂ = FiO₂ × (760 - 47) - PaCO₂ / 0.8, at sea level) and the PaCO₂ minus EtCO₂ gap are calculated from those values. A note marks a draw that may not be at steady state: PaCO₂ less than 20 minutes after a change in rate, VT, inspiratory pressure or mode, PaO₂ less than 5 minutes after a change in FiO₂ or PEEP. In a case, Draw ABG puts the values straight into the case's record table, with no analyzer wait, and the case shows the suggested wait after each change.

Freeze, Pause and the drawn line

Freeze holds the picture while the simulated patient keeps breathing. Pause stops simulation time. Both show the exact last sample. The line is drawn about 0.5 s behind the newest sample so it moves smoothly; every point is an original 50 Hz sample, and no values are filled in to hide a delay.

Fast-forward after each change: the model runs ahead without drawing until the values a change affects settle, for at most one minute of your time, so changes appear faster than in patients. One line above the monitor shows its progress and Stop. Blood gases are drawn only when you press Draw ABG; the Steady or Settling mark beside it says whether the arterial values are still moving. The box in the guide turns the fast-forward off.

What the model does not do

Technical details

Simulation time and elapsed real time appear once the session starts.

No accepted observation.

Ventilation terms

Plain-language meanings of the abbreviations on this page. They describe the terms; they are not settings advice.

Settings

VC · Volume Control
A mode in which the ventilator delivers a set tidal volume with each breath. Airway pressure then rises to whatever level the resistance and compliance of the respiratory system require.
PC · Pressure Control
A mode in which the ventilator holds airway pressure at a set level above PEEP for the inspiratory time. The delivered tidal volume and the flow pattern then depend on resistance and compliance.
VT · Tidal volume
The volume of gas moved into the lungs in one breath, in mL. In VC it is a setting; the VT shown on the monitor is what the model actually delivered.
RR · Respiratory rate
The number of breaths per minute (the Rate setting). Together with Ti or I:E it sets how much time each breath leaves for expiration.
PEEP · Positive end-expiratory pressure
The airway pressure kept above atmospheric pressure at the end of expiration, in cmH₂O. It is the baseline of the pressure waveform between breaths.
FiO₂ · Fraction of inspired oxygen
The share of oxygen in the delivered gas, shown as a percentage; room air is about 21%. It mainly changes oxygenation (PaO₂), not the shape of the waveforms.
Inspiratory pressure (PC)
In Pressure Control, the pressure the ventilator adds above PEEP during inspiration, in cmH₂O.
Ti · Inspiratory time
The time spent in inspiration in each breath, in seconds. Here it is the total inspiratory time: gas delivery plus any inspiratory pause.
I:E · Inspiratory-to-expiratory ratio
The ratio of inspiratory time to expiratory time, written 1:x. A larger x leaves relatively more of each breath for expiration.
Tpause · Inspiratory pause
A short period at the end of inspiration with no gas flow (VC only here). Pressure settles to a plateau during the pause, which is where Pplat is measured.
PC with volume target
A pressure-controlled mode in which the ventilator adjusts the inspiratory pressure after each breath so the delivered tidal volume moves toward the set tidal volume, within a pressure limit. Ventilator makers use different names for this kind of mode.
Rise time
In the pressure modes, the time the ventilator takes to bring airway pressure up to the set level at the start of inspiration, in seconds. A longer rise time gives a softer start and slightly less tidal volume.
Pressure limit
The upper airway pressure of the volume-regulated pressure mode: the inspiratory pressure it chooses stays 5 cmH₂O below this limit, so a stiff lung may not reach the set volume.
Inspiratory flow pattern (VC)
How gas flow is delivered during a Volume Control breath: constant (square) or a decelerating ramp that starts high and falls to zero. The same volume in the same time needs twice the starting flow with a ramp.

Waveforms and pressures

Paw · Airway pressure
The pressure at the airway opening, in cmH₂O, shown in the top waveform.
Flow
Gas flow at the airway, in L/min: positive during inspiration and negative during expiration. Expiratory flow that has not returned to zero when the next breath starts means the lungs had not finished emptying.
Volume · tidal
The volume waveform on the ventilator screen: lung volume above the end-expiratory lung volume of the same breath, in mL. It starts near zero at each breath and peaks at about the tidal volume.
Lung volume · absolute
All the gas in the lungs, in L, including the gas that stays in them between breaths. Tidal volume is the rise and fall of this line within each breath.
Ppeak · Peak inspiratory pressure
The highest airway pressure during inspiration. It is the sum of PEEP, the pressure needed to drive flow through the airways and tube, and the pressure needed to stretch the respiratory system.
Pplat · Plateau pressure
The airway pressure during an inspiratory pause, when flow is zero. Without flow there is no resistive pressure, so it reflects the elastic recoil of the lungs and chest wall at the end of inspiration, including PEEP.
DP · Driving pressure
Plateau pressure minus total PEEP, in cmH₂O. It is the pressure that inflates the respiratory system for one tidal volume, so it needs an inspiratory pause to measure.
Pmean · Mean airway pressure
The average airway pressure over whole breaths, in cmH₂O, here over the last 6 complete breaths.
Intrinsic PEEP · auto-PEEP
Pressure left in the lungs at the end of expiration above the set PEEP, typically because expiration is too short for the lungs to empty fully. Set PEEP plus intrinsic PEEP is total PEEP; here it is a model estimate, not an expiratory-hold measurement.
Inspiratory hold
A maneuver that stops all gas flow at the end of one inspiration for a set time. Airway pressure then settles to the plateau pressure, from which driving pressure, static compliance and (with a constant end-inspiratory flow) resistance are calculated.
ABP · Arterial blood pressure
The pressure in the arteries over each heartbeat, in mmHg, shown as systolic/diastolic with the mean in brackets. Here it is the model value at 50 samples per second.
Capnogram
CO₂ partial pressure at the Y-piece over each breath, in mmHg: near zero while fresh gas flows in, rising during expiration to the end-tidal value.

Calculated values

VTe and VTi · Expiratory and inspiratory tidal volume
VTe is the volume breathed out in the last completed breath, in mL: on the Simulator the expiratory flow summed at the ventilator connection, in Cases the model ventilator value. VTi is the inspiratory flow of the last breath summed at the ventilator connection.
PBW · Predicted body weight
A weight calculated from sex and height: male 50 + 0.91 × (height in cm - 152.4) kg, female 45.5 + 0.91 × (height in cm - 152.4) kg. Tidal volume is shown per kg of it (mL/kg PBW).
Minute ventilation
The volume of gas moved in one minute, in L/min, about tidal volume × respiratory rate.
Resistance · inspiratory
How strongly the airways and tube oppose gas flow, in cmH₂O·s/L. At the same flow, a higher resistance widens the gap between Ppeak and Pplat.
Crs · Respiratory-system compliance
The volume the lungs and chest wall take up per unit of pressure, in mL/cmH₂O, estimated here as VT ÷ (Pplat − total PEEP). A lower compliance means a higher Pplat for the same VT.
Cstat · Static compliance
The ventilator-screen name for respiratory-system compliance measured during an inspiratory pause, in mL/cmH₂O. It is the same quantity as Crs.
Raw · Airway resistance
The ventilator-screen name for the inspiratory resistance estimate, in cmH₂O·s/L.
EtCO₂ · End-tidal CO₂
The CO₂ partial pressure at the end of an exhaled breath, in mmHg. It is usually somewhat lower than PaCO₂.

Arterial blood gases

pH
The acidity of arterial blood. A rise in PaCO₂ lowers pH and a fall in PaCO₂ raises it.
PaO₂ · Arterial oxygen partial pressure
The pressure of oxygen dissolved in arterial blood, in mmHg. It reflects how well oxygen moves from the lungs into the blood; FiO₂ is one of the settings that changes it.
PaCO₂ · Arterial carbon dioxide partial pressure
The pressure of carbon dioxide in arterial blood, in mmHg. More alveolar ventilation lowers it and less ventilation raises it.
HCO₃⁻ · Bicarbonate
The main buffer of the blood, in mmol/L. It changes only slightly when PaCO₂ changes quickly; larger changes take hours to days, so it moves little in a short session.
Measurement methods are in Measurement details and Model information.

Inspiratory pause measurements

Waiting for a qualified pause.

In VC, Tpause is part of total inspiratory time. Increasing it shortens flow delivery and increases the flow required for the same set VT. Requested RR and expiration time are preserved; the native 50 Hz clock quantizes delivered timing.

Pplat is the mean pressure in a qualified, stable zero-flow portion of a completed inspiratory pause. The shaded interval marks the original samples used. Duration and stability affect the estimate; an unavailable value stays blank.

Inspiratory resistance ≈ (end-inspiratory pressure during constant flow − Pplat) / flow in L/s. This effective estimate includes the tube and other resistive components; it is not isolated airway resistance.

Ppeak · flow phase is the maximum pressure in the qualified constant-flow portion, excluding the initial transition. Resistance uses the terminal flow-phase pressure rather than a transient pressure spike.

Respiratory-system compliance (Crs) ≈ same-breath tidal volume / (Pplat − total PEEP). It includes lung and chest wall, not lung compliance alone. Total PEEP is a model estimate, not an expiratory-hold measurement.

A Pplat above 30 cmH₂O adds a short note under the value. In the ARDS Network trial of lower tidal volumes, the lower-volume group kept Pplat at or below 30 cmH₂O, measured after a 0.5 s end-inspiratory pause (N Engl J Med 2000;342:1301-8). The note is for teaching; it is not an alarm.

Results belong to the displayed breath and settings revision. Changing settings clears live results until another breath qualifies. Freeze preserves the original measurement and its time.

Passive model · engineering validity checks, not clinical validation.