Ventilator lessons Lesson 5 of 9
Driving pressure on the ventilator: how to calculate and lower it
Updated
Driving pressure is plateau pressure minus total PEEP. It is the pressure that stretches the lungs and chest wall with each breath, and it equals tidal volume divided by compliance. Smaller breaths lower it. PEEP lowers it only when PEEP opens collapsed lung.
Hands-on lesson · about 4 minLower the tidal volume in ARDSWatch plateau pressure, driving pressure and PaCO₂ respond in the model.How to calculate driving pressure
- Read plateau pressure (Pplat): an inspiratory pause or hold with no flow, in a patient who is not breathing against the ventilator.
- Read total PEEP: an end-expiratory hold. It is the set PEEP plus any auto-PEEP.
- Subtract: driving pressure (DP) = Pplat − total PEEP.
| Formula | DP = Pplat − total PEEP = tidal volume ÷ compliance |
|---|---|
| Example (model ARDS lungs, set VT 480 mL) | 22.7 − 10.3 = 12.4 cmH₂O, and exhaled 460 mL ÷ 12.4 cmH₂O = 37 mL/cmH₂O |
Because DP is volume divided by compliance, the same tidal volume gives a higher driving pressure in stiffer lungs. That is why it is read alongside tidal volume per kg of predicted body weight (PBW), not instead of it.
Lower tidal volume lowers driving pressure
Lowering tidal volume from 480 to 360 mL (8 to 6 mL/kg PBW) lowered driving pressure from 12.4 to 9.1 cmH₂O and plateau pressure from 22.7 to 19.4 cmH₂O. Compliance barely moved (37 and 38 mL/cmH₂O), so DP fell in step with the volume.
The cost is CO₂. At the same rate, PaCO₂ rose from 40.6 to 66.6 mmHg after an hour, and the rate had to go up (adjusting the ventilator from the ABG).
PEEP lowers driving pressure only if it opens lung
Raising PEEP from 5 to 15 cmH₂O at the same tidal volume raised plateau pressure from 17.8 to 27.2 cmH₂O. Driving pressure fell only a little, from 12.5 to 11.7 cmH₂O, because compliance rose from 34.5 to 36.9 mL/cmH₂O as some collapsed lung opened. Oxygenation improved far more: PaO₂ rose from 67 to 107 mmHg and the model's shunt fell from 37% to 19%.
- At the same tidal volume, driving pressure falls after a PEEP increase: compliance rose, so recruitment outweighed overdistension.
- Driving pressure rises after a PEEP increase: compliance fell, so the added PEEP mostly stretched lung that was already open.
Read driving pressure before and after each PEEP change. See the hands-on PEEP lesson for the blood gas side of the same model patient.
Use total PEEP, not set PEEP
With air trapping the set PEEP underestimates the pressure left in the lungs. In model lungs with severe bronchospasm at rate 22/min, Pplat was 18.9 cmH₂O, set PEEP 5 and total PEEP 12.8 cmH₂O. Subtracting set PEEP gives 13.9 cmH₂O; the true driving pressure was 6.1 cmH₂O (auto-PEEP).
What the number means
In a pooled analysis of nine ARDS trials, driving pressure was the ventilation variable most strongly associated with survival: each increase of about 7 cmH₂O was associated with a 41% higher relative risk of death. That is an association: no trial has yet shown that setting the ventilator to a driving pressure target improves survival. Guidelines recommend low tidal volumes (4 to 8 mL/kg PBW) with plateau pressure below 30 cmH₂O in ARDS, and driving pressure is a useful check on top of those.
When the number misleads
- Patient effort: a patient breathing during the pause makes Pplat, and so DP, unreliable.
- Chest wall and abdomen: DP measured at the airway includes the chest wall. A stiff chest wall or a tense abdomen raises it without any extra stretch of the lungs; separating the two needs esophageal pressure.
- Short pause: pressure keeps settling for a moment after flow stops. In this model a 0.5 s pause read 18.2 cmH₂O and a 1 s hold 17.8 cmH₂O.
Key point
Driving pressure is tidal volume seen through compliance. A smaller breath always lowers it. PEEP lowers it only when PEEP opens lung, so check it after every PEEP change.
Try it
- Open the MV Simulator and choose the Stiff lungs patient. The DP value sits beside Pplat.
- Select the VT tile, set a smaller volume and press Confirm. DP falls with the volume.
- Then raise PEEP the same way and compare DP before and after. Small differences from this page are normal.
References
- Amato MBP, Meade MO, Slutsky AS, et al. Driving pressure and survival in the acute respiratory distress syndrome. N Engl J Med 2015;372(8):747-755. doi:10.1056/NEJMsa1410639
- Acute Respiratory Distress Syndrome Network. Ventilation with lower tidal volumes as compared with traditional tidal volumes for acute lung injury and the acute respiratory distress syndrome. N Engl J Med 2000;342(18):1301-1308. doi:10.1056/NEJM200005043421801
- Fan E, Del Sorbo L, Goligher EC, et al. An official American Thoracic Society/European Society of Intensive Care Medicine/Society of Critical Care Medicine clinical practice guideline: mechanical ventilation in adult patients with acute respiratory distress syndrome. Am J Respir Crit Care Med 2017;195(9):1253-1263. doi:10.1164/rccm.201703-0548ST
- Grasselli G, Calfee CS, Camporota L, et al. ESICM guidelines on acute respiratory distress syndrome: definition, phenotyping and respiratory support strategies. Intensive Care Med 2023;49(7):727-759. doi:10.1007/s00134-023-07050-7
- Hess DR. Respiratory mechanics in mechanically ventilated patients. Respir Care 2014;59(11):1773-1794. doi:10.4187/respcare.03410
Saving is off. Marks last for this page only.
Save progress on this device?
Every trace on this page is computed by the TracingsMD ventilator model, not recorded from a patient. For learning only, not for diagnosis or patient care.