Updated

Three pacemaker problems are read from the timing of the spikes. Failure to capture: the pacemaker fires (a spike) but the heart does not respond (no QRS after it). Failure to sense, or undersensing: the pacemaker misses the patient's own beat and fires anyway, so a spike comes too soon after a native beat, sometimes on its T wave. Oversensing works the other way: the pacemaker takes noise or T waves for beats and holds back, leaving pauses with no spikes. The heart model draws capture and loss of capture; undersensing and oversensing are described in the table only.

The strips side by side

Normal ventricular pacing

Normal ventricular pacing ECG strip, lead II, 10 seconds at 25 mm/s and 10 mm/mV
Lead II, 10 seconds (25 mm/s, 10 mm/mV). Every spike is followed at once by a wide QRS: 100% capture at 70/min. Watch normal ventricular pacing on its live strip.

Failure to capture

Failure to capture ECG strip, lead II, 10 seconds at 25 mm/s and 10 mm/mV
Lead II, 10 seconds (25 mm/s, 10 mm/mV). Some spikes have no QRS after them; narrow escape beats fill in, and the pacemaker senses them and waits a full interval. Watch failure to capture on its live strip.

Computed by the TracingsMD heart model: simulated for education, not a patient recording and not for diagnosis.

Here because of a watch alert or a symptom? This site cannot tell what your own heart is doing. Talk to a doctor or nurse about it, and call emergency services for chest pain, fainting or trouble breathing.

Differences at a glance

Scroll the table sideways to read all columns.

FeatureFailure to captureUndersensingOversensing
What you seeA spike with no QRS after it (no P wave after an atrial spike)A spike soon after the patient's own beat, sometimes on its T waveA pause longer than the pacing interval with no spike
What failsThe pacing output does not depolarize the heartThe device does not detect the patient's own beatThe device detects signals that are not beats (muscle noise, T waves, lead noise)
Common causesLead displacement or fracture, a raised threshold (ischemia, fibrosis, drugs, electrolytes), low output, battery depletionLead displacement, a low native signal, a sensing setting that is not sensitive enough, a lead problemMuscle activity, electromagnetic interference, lead fracture, a sensing setting that is too sensitive
Main riskBradycardia or asystole in a patient who depends on pacingA spike on the T wave can rarely start a ventricular arrhythmiaPauses or asystole in a patient who depends on pacing

How to tell them apart

  1. Find every spike and ask what follows it: a QRS right after it is capture; nothing after a spike outside the refractory period is failure to capture.
  2. Find every native beat and ask whether the next spike waited a full pacing interval: a spike too soon after a native beat is undersensing.
  3. Measure every pause: longer than the pacing interval with no spike suggests oversensing or no output.
  4. A spike inside, or right at the start of, a native QRS can be normal fusion or pseudofusion, not a malfunction.
  5. In an asynchronous mode (VOO or DOO, for example with a magnet over the device) the pacemaker does not sense by design, so spikes after native beats are expected.