ECG rhythms Side-by-side comparisons
Pacemaker failure to capture, undersensing and oversensing
Updated
Three pacemaker problems are read from the timing of the spikes. Failure to capture: the pacemaker fires (a spike) but the heart does not respond (no QRS after it). Failure to sense, or undersensing: the pacemaker misses the patient's own beat and fires anyway, so a spike comes too soon after a native beat, sometimes on its T wave. Oversensing works the other way: the pacemaker takes noise or T waves for beats and holds back, leaving pauses with no spikes. The heart model draws capture and loss of capture; undersensing and oversensing are described in the table only.
The strips side by side
Normal ventricular pacing
Failure to capture
Computed by the TracingsMD heart model: simulated for education, not a patient recording and not for diagnosis.
Here because of a watch alert or a symptom? This site cannot tell what your own heart is doing. Talk to a doctor or nurse about it, and call emergency services for chest pain, fainting or trouble breathing.
Differences at a glance
Scroll the table sideways to read all columns.
| Feature | Failure to capture | Undersensing | Oversensing |
|---|---|---|---|
| What you see | A spike with no QRS after it (no P wave after an atrial spike) | A spike soon after the patient's own beat, sometimes on its T wave | A pause longer than the pacing interval with no spike |
| What fails | The pacing output does not depolarize the heart | The device does not detect the patient's own beat | The device detects signals that are not beats (muscle noise, T waves, lead noise) |
| Common causes | Lead displacement or fracture, a raised threshold (ischemia, fibrosis, drugs, electrolytes), low output, battery depletion | Lead displacement, a low native signal, a sensing setting that is not sensitive enough, a lead problem | Muscle activity, electromagnetic interference, lead fracture, a sensing setting that is too sensitive |
| Main risk | Bradycardia or asystole in a patient who depends on pacing | A spike on the T wave can rarely start a ventricular arrhythmia | Pauses or asystole in a patient who depends on pacing |
How to tell them apart
- Find every spike and ask what follows it: a QRS right after it is capture; nothing after a spike outside the refractory period is failure to capture.
- Find every native beat and ask whether the next spike waited a full pacing interval: a spike too soon after a native beat is undersensing.
- Measure every pause: longer than the pacing interval with no spike suggests oversensing or no output.
- A spike inside, or right at the start of, a native QRS can be normal fusion or pseudofusion, not a malfunction.
- In an asynchronous mode (VOO or DOO, for example with a magnet over the device) the pacemaker does not sense by design, so spikes after native beats are expected.
Why it matters
- In a patient who depends on pacing, failure to capture and oversensing can cause sudden bradycardia or asystole: check a pulse and be ready for transcutaneous pacing as your protocol directs.
- Device problems need an interrogation by the device team; a 12-lead ECG and a chest X-ray help show the lead position.