ECG rhythms Pacing

AV-sequential paced rhythm ECG

Also called dual-chamber pacing, AV sequential pacing.

Updated

AV-sequential (dual-chamber) paced rhythm. Two pacing spikes for every beat: the first is followed by a P wave and the second, about 160 ms later, by a wide QRS. The pacemaker drives the atria and ventricles in sequence at 70/min.

Recorded stripHR -

A 30-second recording of the heart model, played in real time. Run live runs the heart model itself with this rhythm.

Computed by the TracingsMD heart model: simulated for education, not a patient recording and not for diagnosis.

Here because of a watch alert or a symptom? This site cannot tell what your own heart is doing. Talk to a doctor or nurse about it, and call emergency services for chest pain, fainting or trouble breathing.

Measured on this recording

Measured on the whole 30-second recording. Rate: the lowest and highest HR the monitor shows as the strip plays (HR is the mean of the latest 8 RR intervals, or of whole repeating cycles such as trigeminy). Intervals: medians of all beats. Run live to watch them update.
Ventricular rate (HR)70/min
PRNot assessable in this rhythm
QRS154 ms
QTNot assessable in this rhythm
QTc (Bazett)Not assessable in this rhythm
QTc (Fridericia)Not assessable in this rhythm

What it is

A rhythm driven by a dual-chamber pacemaker that paces both chambers in sequence: a spike and a paced P wave in the atrium, then, after a programmed AV delay, a second spike and a wide paced QRS in the ventricle. It keeps the atria and ventricles in step when the patient's own AV conduction fails, as in the complete heart block underneath this example. From the usual right ventricular apical lead the QRS looks like left bundle branch block, often with a superior axis, and the T wave points opposite to the QRS; septal or conduction-system pacing gives a different, often narrower QRS. This example paces both chambers on a fixed clock without sensing (DOO); most dual-chamber pacemakers sense and track the patient's own P waves (DDD), so you often see a sinus P wave followed by a ventricular spike instead of two spikes.

How to read this strip

AV-sequential paced rhythm ECG strip, lead II, 10 seconds at 25 mm/s and 10 mm/mV
Lead II, the first 10 seconds of this recording on ECG paper (25 mm/s, 10 mm/mV). Select it to open the full-size image.
Rate70/min, fixed at the programmed pacing rate.
RhythmRegular.
P wavesA paced P wave after each first (atrial) spike.
PRNot assessable as a native PR: the spacing is the programmed AV delay, about 160 ms from the atrial spike to the ventricular spike.
QRSWide (about 150 ms), right after the second spike; negative in V1 (left bundle branch block pattern) and in II (superior axis), as expected from the right ventricular apex.
QTNot assessable: the wide paced QRS lengthens the QT, so it is not read here.

Why patients have it

What to consider

Key point

Two spikes per beat, one before the P wave and one before a wide QRS, is AV-sequential pacing: check that each spike captures its own chamber.

Not to be confused with

Practice with this rhythm

References

  1. Kusumoto FM, Schoenfeld MH, Barrett C, et al. 2018 ACC/AHA/HRS guideline on the evaluation and management of patients with bradycardia and cardiac conduction delay. Circulation 2019;140(8):e382-e482.
  2. Sandau KE, Funk M, Auerbach A, et al. Update to practice standards for electrocardiographic monitoring in hospital settings. Circulation 2017;136(19):e273-e344.

Embed this strip

Paste this code into a web page or a course in Canvas, Moodle or Blackboard. It plays this page's 30-second recording and links back here for the live monitor.