ECG rhythms Pacing
AV-sequential paced rhythm ECG
Also called dual-chamber pacing, AV sequential pacing.
Updated
AV-sequential (dual-chamber) paced rhythm. Two pacing spikes for every beat: the first is followed by a P wave and the second, about 160 ms later, by a wide QRS. The pacemaker drives the atria and ventricles in sequence at 70/min.
A 30-second recording of the heart model, played in real time. Run live runs the heart model itself with this rhythm.
Computed by the TracingsMD heart model: simulated for education, not a patient recording and not for diagnosis.
Here because of a watch alert or a symptom? This site cannot tell what your own heart is doing. Talk to a doctor or nurse about it, and call emergency services for chest pain, fainting or trouble breathing.
Measured on this recording
| Ventricular rate (HR) | 70/min |
|---|---|
| PR | Not assessable in this rhythm |
| QRS | 154 ms |
| QT | Not assessable in this rhythm |
| QTc (Bazett) | Not assessable in this rhythm |
| QTc (Fridericia) | Not assessable in this rhythm |
What it is
A rhythm driven by a dual-chamber pacemaker that paces both chambers in sequence: a spike and a paced P wave in the atrium, then, after a programmed AV delay, a second spike and a wide paced QRS in the ventricle. It keeps the atria and ventricles in step when the patient's own AV conduction fails, as in the complete heart block underneath this example. From the usual right ventricular apical lead the QRS looks like left bundle branch block, often with a superior axis, and the T wave points opposite to the QRS; septal or conduction-system pacing gives a different, often narrower QRS. This example paces both chambers on a fixed clock without sensing (DOO); most dual-chamber pacemakers sense and track the patient's own P waves (DDD), so you often see a sinus P wave followed by a ventricular spike instead of two spikes.
How to read this strip
| Rate | 70/min, fixed at the programmed pacing rate. |
|---|---|
| Rhythm | Regular. |
| P waves | A paced P wave after each first (atrial) spike. |
| PR | Not assessable as a native PR: the spacing is the programmed AV delay, about 160 ms from the atrial spike to the ventricular spike. |
| QRS | Wide (about 150 ms), right after the second spike; negative in V1 (left bundle branch block pattern) and in II (superior axis), as expected from the right ventricular apex. |
| QT | Not assessable: the wide paced QRS lengthens the QT, so it is not read here. |
Why patients have it
- Complete or high-grade AV block, with or without sinus node disease.
- Sinus node dysfunction together with impaired AV conduction.
- Temporary dual-chamber pacing through epicardial wires after cardiac surgery.
What to consider
- Check capture in both chambers: a P wave after each atrial spike and a wide QRS right after each ventricular spike.
- Measure the AV delay from the atrial spike to the ventricular spike; it is programmed, so it should stay the same beat to beat at a fixed rate.
- Paced beats make ST segments and T waves hard to read: ischemia criteria for paced rhythms differ from those for a normal QRS.
- Electrical capture is not mechanical capture: confirm a pulse and perfusion when the rhythm or the patient changes.
- Device questions (rate, mode, battery, leads) go to an interrogation with the patient's device clinic.
Key point
Two spikes per beat, one before the P wave and one before a wide QRS, is AV-sequential pacing: check that each spike captures its own chamber.
Not to be confused with
- Ventricular paced rhythm. Only one spike per beat, before the wide QRS, with no paced P wave.
- Atrial paced rhythm. One spike before each P wave, then a normal PR and a narrow QRS through the patient's own conduction.
- Left bundle branch block. A similar wide QRS after a sinus P wave, with no pacing spikes.
- Ventricular pacing with loss of capture. Some ventricular spikes are not followed by a QRS; here every spike captures.
Side by side: Pacemaker failure to capture, undersensing and oversensing
Practice with this rhythm
References
- Kusumoto FM, Schoenfeld MH, Barrett C, et al. 2018 ACC/AHA/HRS guideline on the evaluation and management of patients with bradycardia and cardiac conduction delay. Circulation 2019;140(8):e382-e482.
- Sandau KE, Funk M, Auerbach A, et al. Update to practice standards for electrocardiographic monitoring in hospital settings. Circulation 2017;136(19):e273-e344.