ECG rhythms Pacing
Ventricular paced rhythm ECG
Also called V-paced rhythm.
Updated
Ventricular paced rhythm. A regular rhythm at 70/min in which every wide QRS starts with a pacing spike: the pacemaker is driving the ventricles.
A 30-second recording of the heart model, played in real time. Run live runs the heart model itself with this rhythm.
Computed by the TracingsMD heart model: simulated for education, not a patient recording and not for diagnosis.
Here because of a watch alert or a symptom? This site cannot tell what your own heart is doing. Talk to a doctor or nurse about it, and call emergency services for chest pain, fainting or trouble breathing.
Measured on this recording
| Ventricular rate (HR) | 70/min |
|---|---|
| PR | Not assessable in this rhythm |
| QRS | 154 ms |
| QT | Not assessable in this rhythm |
| QTc (Bazett) | Not assessable in this rhythm |
| QTc (Fridericia) | Not assessable in this rhythm |
What it is
A rhythm driven by a pacemaker lead in the right ventricle. Each beat starts with a narrow, sharp pacing spike followed at once by a wide QRS: activation spreads from the right ventricle through muscle instead of the conduction system. From the usual right ventricular site the QRS looks like left bundle branch block, often with a superior (left) axis, and the T wave points opposite to the QRS. The rate is the programmed rate when the patient's own rhythm is slower.
How to read this strip
| Rate | 70/min, fixed at the programmed pacing rate. |
|---|---|
| Rhythm | Regular. |
| P waves | None seen here: no atrial activity drives these beats. |
| PR | None: each QRS follows a pacing spike, not a P wave. |
| QRS | Wide (about 150 ms), right after each spike; negative in V1 (left bundle branch block pattern) and in II (superior axis), as expected from the right ventricular apex. |
| QT | Not assessable: the wide paced QRS lengthens the QT, so it is not read here. |
Why patients have it
- Sinus node dysfunction with symptomatic bradycardia or pauses.
- High-grade or complete AV block.
- Atrial fibrillation with a slow ventricular response, or after ablation of the AV node.
- Temporary pacing (transvenous or transcutaneous) for an unstable bradycardia, or after cardiac surgery or myocardial infarction.
What to consider
- Check capture: a QRS should follow every spike that falls outside the refractory period; spikes without a QRS are loss of capture.
- Check sensing: a spike that lands well after the start of the patient's own beat, or on its T wave, suggests undersensing; a spike inside the QRS can be normal pseudofusion. Long pauses without spikes suggest oversensing or failure to output (lead or battery problem).
- Electrical capture is not mechanical capture: confirm a pulse or another perfusion signal, especially with transcutaneous pacing.
- Paced beats make ST segments and T waves hard to read: ischemia criteria for paced rhythms differ from those for a normal QRS.
- Long-term right ventricular pacing can weaken the left ventricle; new heart failure symptoms in a frequently paced patient deserve evaluation.
- Device questions (rate, mode, battery, leads) go to an interrogation with the patient's device clinic.
Key point
A pacing spike before every wide QRS is a ventricular paced rhythm: check that every spike captures, that the pacemaker senses the patient's own beats, and that the patient has a pulse.
Not to be confused with
- Left bundle branch block. A similar wide QRS, but each beat starts with a sinus P wave and there is no pacing spike.
- Ventricular pacing with loss of capture. Some spikes are not followed by a QRS; here every spike captures.
- Idioventricular rhythm. A slow wide escape rhythm with no pacing spikes.
- Monomorphic ventricular tachycardia. Wide and regular but fast (above 100/min) and without pacing spikes.
Side by side: Pacemaker failure to capture, undersensing and oversensing
Practice with this rhythm