ECG rhythms Pacing
Atrial paced rhythm ECG
Also called atrial pacing.
Updated
Atrial paced rhythm. A regular rhythm at 70/min in which every P wave follows a pacing spike and conducts with a normal PR to a narrow QRS: the pacemaker is driving the atria.
A 30-second recording of the heart model, played in real time. Run live runs the heart model itself with this rhythm.
Computed by the TracingsMD heart model: simulated for education, not a patient recording and not for diagnosis.
Here because of a watch alert or a symptom? This site cannot tell what your own heart is doing. Talk to a doctor or nurse about it, and call emergency services for chest pain, fainting or trouble breathing.
Measured on this recording
| Ventricular rate (HR) | 70/min |
|---|---|
| PR | 158 ms |
| QRS | 84 ms |
| QT | 359 ms |
| QTc (Bazett) | 387 ms |
| QTc (Fridericia) | 378 ms |
What it is
A rhythm driven by a pacemaker lead in the right atrium. Each beat starts with a small pacing spike followed by a paced P wave, then a normal PR interval and a narrow QRS: the impulse reaches the ventricles through the patient's own AV node and conduction system. Atrial pacing is used when the sinus node is too slow but AV conduction is normal. The spike can be very small, especially from a bipolar lead, and may be hard to see in some leads. This example paces at a fixed rate without sensing (AOO); most implanted pacemakers sense the patient's own P waves and pace only when the sinus rate falls below the programmed rate.
How to read this strip
| Rate | 70/min, fixed at the programmed pacing rate. |
|---|---|
| Rhythm | Regular. |
| P waves | One paced P wave right after each spike; no sinus P waves are seen. |
| PR | Normal and constant (about 160 ms), measured from the paced P wave. |
| QRS | Narrow (about 85 ms): the ventricles are activated through the normal conduction system. |
| QT | Normal. |
Why patients have it
- Sinus node dysfunction (sick sinus syndrome) with symptomatic bradycardia or pauses, when AV conduction is normal.
- A dual-chamber pacemaker pacing the atrium whenever the sinus rate falls below its lower rate while the AV node still conducts.
- Temporary atrial pacing through epicardial wires after cardiac surgery.
What to consider
- Check atrial capture: a P wave should follow each spike, and each P wave should conduct to a QRS.
- Atrial pacing depends on AV conduction: if AV block develops, the paced P waves are no longer followed by a QRS and the ventricles need their own pacing.
- Atrial pacing cannot capture fibrillating atria: in atrial fibrillation only ventricular pacing can prevent a slow ventricular rate.
- Electrical capture is not mechanical capture: confirm a pulse and perfusion when the rhythm or the patient changes.
- Device questions (rate, mode, battery, leads) go to an interrogation with the patient's device clinic.
Key point
A spike before each P wave, a normal PR and a narrow QRS is atrial pacing: the pacemaker drives the atria and the patient's own AV node carries each beat to the ventricles.
Not to be confused with
- Normal sinus rhythm. The same narrow QRS after a normal PR, but no spike before the P wave and the rate varies a little with breathing and activity.
- Ventricular paced rhythm. The spike comes right before a wide QRS, with no P wave driving the beat.
- AV-sequential paced rhythm. Two spikes per beat, one before the P wave and one before a wide QRS.
- Atrial pacing with loss of capture (simulator). Some atrial spikes are not followed by a P wave, so those beats are missing or replaced by an escape beat.
Side by side: Pacemaker failure to capture, undersensing and oversensing
Practice with this rhythm
References
- Kusumoto FM, Schoenfeld MH, Barrett C, et al. 2018 ACC/AHA/HRS guideline on the evaluation and management of patients with bradycardia and cardiac conduction delay. Circulation 2019;140(8):e382-e482.
- Sandau KE, Funk M, Auerbach A, et al. Update to practice standards for electrocardiographic monitoring in hospital settings. Circulation 2017;136(19):e273-e344.