ECG Basics Part 1: Start here Lesson 3 of 22

What an ECG records

An ECG is a graph of voltage over time, measured between electrodes on the skin. Learn what an electrode, a lead and the baseline are before reading any wave.

Electricity, not pumping

An ECG (electrocardiogram) plots voltage on the vertical axis against time on the horizontal axis. Each contraction follows its electrical signal a moment later, but the ECG only records the electricity. It cannot show whether blood is actually moving: it gives no pulse and no blood pressure (see Lesson 22: Arrest rhythms, pacing and artifact).

Electrodes and leads

A bedside monitor usually uses three electrodes (right arm, left arm, left leg) or five (adding a right-leg neutral and one chest electrode). A diagnostic 12-lead ECG uses ten electrodes to make twelve leads (Lesson 14: The 12 leads and what they see).

Limb electrodes and the direction of lead IIFront view of a torso. The heart lies behind the sternum with its apex toward the left fifth intercostal space at the mid-clavicular line. Electrodes: RA on the right shoulder, LA on the left shoulder, LL on the left lower torso, and RL on the right lower torso as the neutral electrode. Lead II looks from RA (negative) to LL (positive), about plus 60 degrees. Inset: a wave travelling toward the positive electrode gives an upward deflection, away gives a downward deflection, and across gives a small or biphasic deflection.RLRALALLRL−+(neutral)Sternumheart lies behind itHeartapex: left 5th intercostalspace, mid-clavicular lineMid-clavicular lineLead II: RA (−) → LL (+), about +60°How a lead sees a wave+Toward + → upward deflection+Away from + → downward+Across → small or biphasic
Monitoring electrode positions on the torso; for a diagnostic 12-lead ECG the limb electrodes go on the limbs. Lead II compares LL (+) with RA (−), so it looks along the heart from upper right to lower left, about +60°. Inset: arrows show deflection direction only, not ECG waveforms.
ElectrodeAHA colourIEC colour
Right arm (RA / R)WhiteRed
Left arm (LA / L)BlackYellow
Left leg (LL / F)RedGreen
Right leg (RL / N)GreenBlack

Colour codes differ between regions, so always read the letters printed on the clip.

One rule for up and down

Lead II measures from the right arm (negative) to the left leg (positive). The normal impulse runs from the SA node at the top right of the heart toward the ventricles at the bottom left, roughly toward lead II's positive electrode. So in lead II the P wave, QRS and T wave are all upright, which is one reason monitors and rhythm strips so often show lead II.

Lead aVR looks at the heart from the right shoulder. The same beats, recorded at the same moment, travel away from it, so they point downward:

The grid in one minute

At the standard speed of 25 mm/s, one small box is 0.04 s and one large box is 0.20 s, so five large boxes make one second. Vertically, 10 mm is 1 mV. The full details are in Lesson 7: The grid and calibration.

The baseline

The flat line between the end of one T wave and the next P wave (the TP segment) is the baseline. Other parts of the trace are judged against it.

Signals that are not from the heart

Movement, shivering, loose electrodes and mains interference all add artifact, which can hide the rhythm or imitate a dangerous one (Lesson 22: Arrest rhythms, pacing and artifact). Look at the patient, not only at the screen.

Key points

  • The ECG records electrical activity only; it cannot show a pulse.
  • Electrodes are pads; leads are viewpoints made from them.
  • Toward the positive electrode = up; away = down.
  • In lead II the normal P, QRS and T are upright.
Check yourself
  1. A wave of activation travels toward a lead's positive electrode. Does the trace go up or down?

    Answer: up.

  2. Which electrode is positive in lead II?

    Answer: the left leg electrode.

  3. Can a normal-looking ECG prove that the patient has a pulse?

    Answer: no. The ECG shows electrical activity only.

See it move in the live simulatorNormal sinus rhythm

To see all twelve viewpoints at once, open normal sinus rhythm in the 12-lead view.

References

  1. Kligfield P, et al. Recommendations for the standardization and interpretation of the electrocardiogram, Part I: the electrocardiogram and its technology. Circulation 2007;115:1306–1324.
  2. Hall JE, Hall ME. Guyton and Hall Textbook of Medical Physiology. 14th ed. Elsevier; 2021.