ECG Basics · Lesson 4 · Reading the paper

The grid and calibration

Every measurement on an ECG relies on two settings: paper speed and gain. Check them first, then the grid turns into a ruler.

Time runs left to right

At the standard speed of 25 mm/s:

Voltage runs bottom to top

At the standard gain of 10 mm/mV, one small box = 0.1 mV and two large boxes = 1 mV. Many machines print a calibration pulse at the start: a rectangle 10 mm tall (1 mV) and 5 mm wide.

When the settings change

SettingEffectWhy it is used
50 mm/s1 small box = 20 ms; everything looks twice as wideSpreading out fast rhythms or fine detail
12.5 mm/s1 small box = 80 ms; complexes look crowdedLong rhythm overviews
5 mm/mVHalf height; calibration pulse 5 mm tallVery tall QRS complexes
20 mm/mVDouble heightSmall complexes
If a tracing looks oddly fast, slow, tall or flat, check the speed and calibration labels before calling anything abnormal.

Here is a real-scale strip from the engine: count the large boxes between two R waves and you already have the rate (next lessons).

The live simulator lets you switch between 12.5, 25 and 50 mm/s and between 5, 10 and 20 mm/mV while the same rhythm keeps running. Open Settings → Display to try it.