ECG Practice ECG quiz

ECG quiz questions with answers

Updated

24 multiple-choice questions on ECG paper, intervals, rate, conduction, axis, ST elevation, QT and arrest rhythms, at three levels. Each answer explains why every choice is right or wrong and links the lesson that covers it.

Prefer to be scored? Take the same questions in the interactive ECG theory practice, which tracks what you got right and brings back what you missed.

Beginner (9)

1. At 25 mm/s, how much time is one 1 mm box?

  1. 40 ms
  2. 200 ms
  3. 1 second
  4. 10 ms
Show the answerHide the answer

Answer A: 40 ms. Time = distance ÷ speed: 1 mm ÷ 25 mm/s = 0.04 s, or 40 ms. Always check the displayed paper speed first.

  • A. 40 ms: Correct: 1 mm ÷ 25 mm/s = 0.04 s.
  • B. 200 ms: 200 ms is one large (5 mm) box, not one small box.
  • C. 1 second: One second is 25 mm, or five large boxes, at this speed.
  • D. 10 ms: 10 ms per mm would need a paper speed of 100 mm/s.

References:

  1. Kligfield P, Gettes LS, Bailey JJ, et al. Recommendations for the standardization and interpretation of the electrocardiogram, Part I: the electrocardiogram and its technology. Circulation 2007;115(10):1306-1324. doi:10.1161/CIRCULATIONAHA.106.180200

2. At 10 mm/mV, what voltage does one 1 mm box represent?

  1. 0.1 mV
  2. 1 mV
  3. 10 mV
  4. 0.5 mV
Show the answerHide the answer

Answer A: 0.1 mV. Voltage = distance ÷ gain: 1 mm ÷ 10 mm/mV = 0.1 mV. Changing gain changes the drawing scale, not the underlying samples.

  • A. 0.1 mV: Correct: 1 mm ÷ 10 mm/mV = 0.1 mV.
  • B. 1 mV: 1 mV is 10 mm, the height of the standard calibration pulse.
  • C. 10 mV: 10 mV would be 100 mm tall at this gain.
  • D. 0.5 mV: 0.5 mV is one large (5 mm) box, not one small box.

References:

  1. Kligfield P, Gettes LS, Bailey JJ, et al. Recommendations for the standardization and interpretation of the electrocardiogram, Part I: the electrocardiogram and its technology. Circulation 2007;115(10):1306-1324. doi:10.1161/CIRCULATIONAHA.106.180200

3. Which adult QRS width is the wide-complex boundary in the AHA tachyarrhythmia algorithm?

  1. 120 ms
  2. 40 ms
  3. 80 ms
  4. 200 ms
Show the answerHide the answer

Answer A: 120 ms. The AHA adult algorithm uses QRS ≥ 0.12 s (120 ms) as wide. Width organizes the assessment; it does not prove a ventricular origin.

  • A. 120 ms: Correct: the algorithm calls a QRS of 0.12 s (120 ms) or more wide.
  • B. 40 ms: 40 ms is a single small box; no QRS boundary uses it.
  • C. 80 ms: 80 ms is within the normal QRS range, not the wide boundary.
  • D. 200 ms: 200 ms is the upper limit of a normal PR interval, not a QRS boundary.

References:

  1. American Heart Association. Adult tachyarrhythmia with a pulse algorithm, 2025. Source
  2. Wigginton JG, Agarwal S, Bartos JA, et al. Part 9: adult advanced life support: 2025 American Heart Association guidelines for cardiopulmonary resuscitation and emergency cardiovascular care. Circulation 2025;152(16 Suppl 2):S538-S577. doi:10.1161/CIR.0000000000001376

4. A rhythm is regular at 25 mm/s. R waves are 4 large boxes apart. What is the heart rate?

  1. 60/min
  2. 75/min
  3. 100/min
  4. 150/min
Show the answerHide the answer

Answer B: 75/min. Use the 300 rule: 300 ÷ 4 = 75/min. It works because one large box is 0.2 s, so 300 large boxes pass in one minute at 25 mm/s. For more precision, count small boxes and divide 1500 by that number.

  • A. 60/min: 60/min would need 5 large boxes between R waves.
  • B. 75/min: Correct: 300 ÷ 4 = 75/min.
  • C. 100/min: 100/min would need 3 large boxes between R waves.
  • D. 150/min: 150/min would need 2 large boxes between R waves.

References:

  1. Kligfield P, Gettes LS, Bailey JJ, et al. Recommendations for the standardization and interpretation of the electrocardiogram, Part I: the electrocardiogram and its technology. Circulation 2007;115(10):1306-1324. doi:10.1161/CIRCULATIONAHA.106.180200

5. The rhythm is irregularly irregular, as in atrial fibrillation. Which rate method is most reliable?

  1. Apply the 300 rule to one RR interval
  2. Apply the 1500 rule to the shortest RR interval
  3. Count QRS complexes in 6 seconds and multiply by 10
  4. Count P waves in one large box
Show the answerHide the answer

Answer C: Count QRS complexes in 6 seconds and multiply by 10. Box rules use a single RR interval, so they only work when every interval is the same. With an irregular rhythm, count the QRS complexes over a set time instead. Six seconds is 30 large boxes at 25 mm/s; on a 10-second strip, multiply the count by 6.

  • A. Apply the 300 rule to one RR interval: One RR interval can be far shorter or longer than the average in an irregular rhythm.
  • B. Apply the 1500 rule to the shortest RR interval: The shortest interval gives the fastest instantaneous rate, which overestimates the average.
  • C. Count QRS complexes in 6 seconds and multiply by 10: Correct: counting beats over 6 seconds averages out the irregular spacing.
  • D. Count P waves in one large box: P waves give the atrial rate, and AF has no discrete P waves to count.

References:

  1. Kligfield P, Gettes LS, Bailey JJ, et al. Recommendations for the standardization and interpretation of the electrocardiogram, Part I: the electrocardiogram and its technology. Circulation 2007;115(10):1306-1324. doi:10.1161/CIRCULATIONAHA.106.180200

6. What is the usual adult PR interval range?

  1. 200-280 ms
  2. 80-120 ms
  3. 350-450 ms
  4. 120-200 ms
Show the answerHide the answer

Answer D: 120-200 ms. Most references use 120-200 ms in adults, measured from the start of the P wave to the start of the QRS. A PR above 200 ms with every P conducted is first-degree AV block. A short PR can suggest pre-excitation or a low atrial or junctional origin.

  • A. 200-280 ms: A constant PR above 200 ms is first-degree AV block, not normal.
  • B. 80-120 ms: 80-120 ms overlaps the QRS range; a PR under 120 ms is short.
  • C. 350-450 ms: 350-450 ms is in the range of a QT interval, not a PR interval.
  • D. 120-200 ms: Correct: 120-200 ms (3 to 5 small boxes at 25 mm/s).

References:

  1. Kusumoto FM, Schoenfeld MH, Barrett C, et al. 2018 ACC/AHA/HRS guideline on the evaluation and management of patients with bradycardia and cardiac conduction delay. Circulation 2019;140(8):e382-e482. doi:10.1161/CIR.0000000000000628

7. In adults, what QRS duration does AHA Part III require for complete right or left bundle branch block?

  1. ≥ 120 ms
  2. ≥ 100 ms
  3. ≥ 160 ms
  4. ≥ 80 ms
Show the answerHide the answer

Answer A: ≥ 120 ms. AHA Part III defines complete RBBB and complete LBBB in adults with a QRS of 120 ms or more. A similar shape with a QRS of 110-119 ms is called incomplete. Width alone does not name the block; the QRS shape in V1, V6 and lead I does.

  • A. ≥ 120 ms: Correct: 120 ms (3 small boxes) or more in adults.
  • B. ≥ 100 ms: 100 ms is still within the normal adult range.
  • C. ≥ 160 ms: 160 ms is wider than required; complete block starts at 120 ms.
  • D. ≥ 80 ms: 80 ms is a normal QRS width.

References:

  1. Surawicz B, Childers R, Deal BJ, et al. Recommendations for the standardization and interpretation of the electrocardiogram, Part III: intraventricular conduction disturbances. J Am Coll Cardiol 2009;53(11):976-981. doi:10.1016/j.jacc.2008.12.013

8. Why is the QT interval corrected for heart rate (QTc)?

  1. Correction removes U waves from the measurement
  2. QT normally shortens as the rate rises, so raw QT values are hard to compare across rates
  3. QTc is the same as the PR interval at a rate of 60/min
  4. Machines always measure the raw QT wrongly
Show the answerHide the answer

Answer B: QT normally shortens as the rate rises, so raw QT values are hard to compare across rates. Repolarization is shorter at faster rates, so the same raw QT can be normal at one rate and long at another. QTc estimates the QT at a rate of 60/min. Bazett's formula (QT ÷ √RR) is common but over-corrects at fast rates; AHA Part IV recommends linear regression formulas instead.

  • A. Correction removes U waves from the measurement: Rate correction does not remove U waves; deciding where the T wave ends is a separate step.
  • B. QT normally shortens as the rate rises, so raw QT values are hard to compare across rates: Correct: QT varies with rate, so it is adjusted to a standard rate before comparing.
  • C. QTc is the same as the PR interval at a rate of 60/min: QTc relates to QT only; it has nothing to do with the PR interval.
  • D. Machines always measure the raw QT wrongly: Automated QT measurement can err, but that is not why rate correction exists.

References:

  1. Rautaharju PM, Surawicz B, Gettes LS, et al. Recommendations for the standardization and interpretation of the electrocardiogram, Part IV: the ST segment, T and U waves, and the QT interval. J Am Coll Cardiol 2009;53(11):982-991. doi:10.1016/j.jacc.2008.12.014

9. The net QRS is positive in lead I and negative in aVF. What can you say about the frontal axis?

  1. It is right axis deviation
  2. It is an extreme (northwest) axis
  3. It lies between 0° and −90°; lead II separates normal from left axis deviation
  4. It is normal in every case
Show the answerHide the answer

Answer C: It lies between 0° and −90°; lead II separates normal from left axis deviation. Lead I positive points the axis to the left; aVF negative points it upward, so the axis is between 0° and −90°. The normal range is about −30° to +90°. If lead II is mainly positive the axis is still normal; if lead II is mainly negative it is left axis deviation.

  • A. It is right axis deviation: Right axis deviation needs a negative lead I and positive aVF.
  • B. It is an extreme (northwest) axis: An extreme axis needs both lead I and aVF negative.
  • C. It lies between 0° and −90°; lead II separates normal from left axis deviation: Correct: this is the left upper quadrant, and lead II decides whether it is past −30°.
  • D. It is normal in every case: Axes from −30° to −90° in this quadrant are left axis deviation, so it is not always normal.

References:

  1. Kligfield P, Gettes LS, Bailey JJ, et al. Recommendations for the standardization and interpretation of the electrocardiogram, Part I: the electrocardiogram and its technology. Circulation 2007;115(10):1306-1324. doi:10.1161/CIRCULATIONAHA.106.180200
  2. Surawicz B, Childers R, Deal BJ, et al. Recommendations for the standardization and interpretation of the electrocardiogram, Part III: intraventricular conduction disturbances. J Am Coll Cardiol 2009;53(11):976-981. doi:10.1016/j.jacc.2008.12.013

Intermediate (9)

10. What can a fast narrow QRS establish by itself?

  1. A description of ventricular activation, not a complete rhythm diagnosis
  2. The exact SVT mechanism
  3. That the patient has a pulse
  4. That it is always sinus tachycardia
Show the answerHide the answer

Answer A: A description of ventricular activation, not a complete rhythm diagnosis. A narrow fast rhythm still needs regularity, atrial activity and clinical context. Different supraventricular mechanisms can produce similar monitor strips.

  • A. A description of ventricular activation, not a complete rhythm diagnosis: Correct: a narrow QRS tells you the ventricles were activated quickly, usually through the normal conduction system.
  • B. The exact SVT mechanism: AVNRT, AVRT, atrial tachycardia and flutter can all look alike; width alone cannot separate them.
  • C. That the patient has a pulse: An ECG cannot show a pulse; that needs a clinical check.
  • D. That it is always sinus tachycardia: Sinus tachycardia is one possibility among several narrow-complex rhythms.

References:

  1. Page RL, Joglar JA, Caldwell MA, et al. 2015 ACC/AHA/HRS guideline for the management of adult patients with supraventricular tachycardia. Circulation 2016;133(14):e506-e574. doi:10.1161/CIR.0000000000000311
  2. American Heart Association. Adult tachyarrhythmia with a pulse algorithm, 2025. Source

11. In a model with an atrial rate of 300/min and fixed 2:1 conduction, what ventricular rate follows?

  1. 150/min
  2. 300/min
  3. 600/min
  4. 75/min
Show the answerHide the answer

Answer A: 150/min. Two atrial cycles produce one ventricular beat: 300 ÷ 2 = 150/min. The ratio is the teaching point; a real rate near 150 alone does not diagnose flutter.

  • A. 150/min: Correct: 300 ÷ 2 = 150/min.
  • B. 300/min: 300/min would mean 1:1 conduction, with every atrial cycle conducted.
  • C. 600/min: The ventricles cannot beat faster than the atria they follow here.
  • D. 75/min: 75/min would be 4:1 conduction.

References:

  1. Page RL, Joglar JA, Caldwell MA, et al. 2015 ACC/AHA/HRS guideline for the management of adult patients with supraventricular tachycardia. Circulation 2016;133(14):e506-e574. doi:10.1161/CIR.0000000000000311

12. Can AF have a regular ventricular rhythm in some circumstances?

  1. Yes, for example with AV block or ventricular pacing
  2. No, every AF strip must have irregular RR intervals
  3. Only if every QRS is narrow
  4. Only if the atrial rate is below 60/min
Show the answerHide the answer

Answer A: Yes, for example with AV block or ventricular pacing. AF usually has irregular RR spacing, but AV block or a paced ventricular rhythm can make the QRS spacing regular. Assess atrial activity too.

  • A. Yes, for example with AV block or ventricular pacing: Correct: with complete AV block an escape rhythm, or a ventricular pacemaker, sets a regular QRS rhythm.
  • B. No, every AF strip must have irregular RR intervals: Irregular RR spacing is typical, but it is not absolute.
  • C. Only if every QRS is narrow: A regular escape or paced rhythm is often wide, so a narrow QRS is not required.
  • D. Only if the atrial rate is below 60/min: AF has no organized atrial rate of this kind; atrial activity is rapid and disorganized.

References:

  1. Joglar JA, Chung MK, Armbruster AL, et al. 2023 ACC/AHA/ACCP/HRS guideline for the diagnosis and management of atrial fibrillation. Circulation 2024;149(1):e1-e156. Table 4. doi:10.1161/CIR.0000000000001193

13. Can this simulator's organized ECG prove that blood is circulating?

  1. No; ECG activity does not establish a pulse or perfusion
  2. Yes, if QRS complexes repeat
  3. Yes, if the displayed rate is normal
  4. Yes, if a P wave precedes each QRS
Show the answerHide the answer

Answer A: No; ECG activity does not establish a pulse or perfusion. An ECG records electrical activity. Pulse and perfusion require clinical assessment; this model does not simulate circulation.

  • A. No; ECG activity does not establish a pulse or perfusion: Correct: electrical activity and circulation are separate questions.
  • B. Yes, if QRS complexes repeat: Repeating QRS complexes can occur without a pulse, as in PEA.
  • C. Yes, if the displayed rate is normal: A normal displayed rate says nothing about cardiac output.
  • D. Yes, if a P wave precedes each QRS: Normal P-QRS order is still electrical information only.

References:

  1. Wigginton JG, Agarwal S, Bartos JA, et al. Part 9: adult advanced life support: 2025 American Heart Association guidelines for cardiopulmonary resuscitation and emergency cardiovascular care. Circulation 2025;152(16 Suppl 2):S538-S577. doi:10.1161/CIR.0000000000001376
  2. American Heart Association. Adult cardiac arrest algorithm, 2025. Source

14. What does a pacing spike show by itself?

  1. Electrical output, without proving capture or circulation
  2. Successful mechanical capture
  3. Return of spontaneous circulation
  4. A normal sinus beat
Show the answerHide the answer

Answer A: Electrical output, without proving capture or circulation. Output, electrical depolarization and evidence of circulation are separate observations. A spike alone establishes neither electrical nor mechanical capture. This is a concept exercise, not an arrest treatment exercise.

  • A. Electrical output, without proving capture or circulation: Correct: a spike shows only that the pacemaker delivered output.
  • B. Successful mechanical capture: Mechanical capture needs a palpable pulse or other evidence of perfusion.
  • C. Return of spontaneous circulation: ROSC is a clinical finding; a spike on the monitor cannot show it.
  • D. A normal sinus beat: A paced spike comes from a device, not from the sinus node.

References:

  1. Wigginton JG, Agarwal S, Bartos JA, et al. Part 9: adult advanced life support: 2025 American Heart Association guidelines for cardiopulmonary resuscitation and emergency cardiovascular care. Circulation 2025;152(16 Suppl 2):S538-S577. doi:10.1161/CIR.0000000000001376

15. A noisy tracing resembles a dangerous rhythm. What is the sound interpretation principle?

  1. Consider signal quality, other leads and the patient's condition
  2. Any noisy tracing proves VF
  3. A filter can guarantee the correct diagnosis
  4. Electrical artifact proves there is no pulse
Show the answerHide the answer

Answer A: Consider signal quality, other leads and the patient's condition. Artifact can obscure the underlying rhythm. The AHA notes that chest compressions create ECG artifact; signal interpretation must be related to the clinical situation.

  • A. Consider signal quality, other leads and the patient's condition: Correct: check the leads, other channels and the patient before naming the rhythm.
  • B. Any noisy tracing proves VF: Movement, tremor or loose leads can mimic VF.
  • C. A filter can guarantee the correct diagnosis: Filters change the drawing and can hide or distort real signals.
  • D. Electrical artifact proves there is no pulse: Artifact is unrelated to whether a pulse is present.

References:

  1. Wigginton JG, Agarwal S, Bartos JA, et al. Part 9: adult advanced life support: 2025 American Heart Association guidelines for cardiopulmonary resuscitation and emergency cardiovascular care. Circulation 2025;152(16 Suppl 2):S538-S577. doi:10.1161/CIR.0000000000001376

16. The QRS is 140 ms. V1 shows an rsR′ pattern, and leads I and V6 show a wide, slurred S wave. Which pattern fits best?

  1. Left bundle branch block
  2. Left anterior fascicular block
  3. Normal intraventricular conduction
  4. Right bundle branch block
Show the answerHide the answer

Answer D: Right bundle branch block. In RBBB the right ventricle is activated late, which gives a terminal R′ in V1 and a wide terminal S wave in I and V6. In LBBB, V1 shows a deep QS or rS, and I, aVL and V5-V6 show a broad, notched or slurred R wave without septal q waves. Look at V1 first, then V6.

  • A. Left bundle branch block: LBBB gives a deep QS or rS in V1 and a broad R in V6, the opposite of this pattern.
  • B. Left anterior fascicular block: Fascicular block mainly shifts the axis and does not widen the QRS to 140 ms by itself.
  • C. Normal intraventricular conduction: A QRS of 140 ms is wide, so conduction is not normal.
  • D. Right bundle branch block: Correct: wide QRS, rsR′ in V1 and a wide S in I and V6 are the RBBB pattern.

References:

  1. Surawicz B, Childers R, Deal BJ, et al. Recommendations for the standardization and interpretation of the electrocardiogram, Part III: intraventricular conduction disturbances. J Am Coll Cardiol 2009;53(11):976-981. doi:10.1016/j.jacc.2008.12.013

17. Which finding best separates Mobitz II from Mobitz I second-degree AV block?

  1. QRS width alone
  2. A constant PR before a P wave that suddenly fails to conduct
  3. PR lengthening beat by beat before the dropped beat
  4. A ventricular rate below 40/min
Show the answerHide the answer

Answer B: A constant PR before a P wave that suddenly fails to conduct. In Mobitz I the PR lengthens until a P wave fails to conduct, and the PR after the pause is shorter. In Mobitz II the PR stays the same before and after the nonconducted P wave. With 2:1 conduction the strip alone cannot tell the two apart.

  • A. QRS width alone: Mobitz II often has a wide QRS, but it can be narrow; width is supportive, not defining.
  • B. A constant PR before a P wave that suddenly fails to conduct: Correct: a fixed PR with a sudden dropped beat is Mobitz II.
  • C. PR lengthening beat by beat before the dropped beat: Progressive PR lengthening is Mobitz I (Wenckebach).
  • D. A ventricular rate below 40/min: Rate does not define the type of block; both types can give slow or near-normal rates.

References:

  1. Kusumoto FM, Schoenfeld MH, Barrett C, et al. 2018 ACC/AHA/HRS guideline on the evaluation and management of patients with bradycardia and cardiac conduction delay. Circulation 2019;140(8):e382-e482. doi:10.1161/CIR.0000000000000628

18. What defines third-degree (complete) AV block on the ECG?

  1. Every P wave is followed by a QRS, with a PR above 200 ms
  2. P waves and QRS complexes march out independently, with no P waves conducting
  3. The PR lengthens until one P wave is not conducted
  4. An irregularly irregular rhythm with no visible P waves
Show the answerHide the answer

Answer B: P waves and QRS complexes march out independently, with no P waves conducting. In complete AV block no atrial impulse reaches the ventricles, so P waves and QRS complexes are dissociated. The atrial rate is faster than the ventricular escape rate. A narrow escape suggests a junctional origin; a wide escape suggests a ventricular origin.

  • A. Every P wave is followed by a QRS, with a PR above 200 ms: Every P conducting with a long PR is first-degree AV block.
  • B. P waves and QRS complexes march out independently, with no P waves conducting: Correct: AV dissociation with no conducted P waves, and the atria faster than the escape rhythm.
  • C. The PR lengthens until one P wave is not conducted: This is Mobitz I second-degree block; some P waves still conduct.
  • D. An irregularly irregular rhythm with no visible P waves: This describes atrial fibrillation, not AV block.

References:

  1. Kusumoto FM, Schoenfeld MH, Barrett C, et al. 2018 ACC/AHA/HRS guideline on the evaluation and management of patients with bradycardia and cardiac conduction delay. Circulation 2019;140(8):e382-e482. doi:10.1161/CIR.0000000000000628

Advanced (6)

19. What makes an organized electrical rhythm PEA?

  1. The absence of a detectable pulse in the clinical context
  2. A unique QRS shape
  3. An ECG heart rate of zero
  4. Any noisy baseline
Show the answerHide the answer

Answer A: The absence of a detectable pulse in the clinical context. PEA is a clinical combination of electrical activity and absent pulse, not one special ECG shape. It cannot be confirmed from these model traces alone.

  • A. The absence of a detectable pulse in the clinical context: Correct: PEA is organized electrical activity with no pulse on clinical assessment.
  • B. A unique QRS shape: PEA can show almost any organized rhythm; no single QRS shape defines it.
  • C. An ECG heart rate of zero: PEA has electrical activity by definition; an ECG rate of zero is asystole.
  • D. Any noisy baseline: Noise is a signal-quality problem, not a rhythm diagnosis.

References:

  1. American Heart Association. Adult cardiac arrest algorithm, 2025. Source
  2. Wigginton JG, Agarwal S, Bartos JA, et al. Part 9: adult advanced life support: 2025 American Heart Association guidelines for cardiopulmonary resuscitation and emergency cardiovascular care. Circulation 2025;152(16 Suppl 2):S538-S577. doi:10.1161/CIR.0000000000001376

20. Which QTc thresholds does AHA Part IV give for a prolonged QT in adults?

  1. ≥ 400 ms for both sexes
  2. ≥ 500 ms in men and ≥ 520 ms in women
  3. ≥ 450 ms in men and ≥ 460 ms in women
  4. ≥ 390 ms for both sexes
Show the answerHide the answer

Answer C: ≥ 450 ms in men and ≥ 460 ms in women. AHA Part IV defines a prolonged QTc as 450 ms or more in men and 460 ms or more in women. A QTc above 500 ms is widely treated as a marker of higher torsades risk. Check for QT-prolonging drugs and low potassium or magnesium.

  • A. ≥ 400 ms for both sexes: 400 ms is a normal QTc for most adults.
  • B. ≥ 500 ms in men and ≥ 520 ms in women: These values are too high; prolongation starts below 500 ms.
  • C. ≥ 450 ms in men and ≥ 460 ms in women: Correct: ≥ 450 ms in men and ≥ 460 ms in women.
  • D. ≥ 390 ms for both sexes: 390 ms or less is the Part IV cut-off for a short QT, not a long one.

References:

  1. Rautaharju PM, Surawicz B, Gettes LS, et al. Recommendations for the standardization and interpretation of the electrocardiogram, Part IV: the ST segment, T and U waves, and the QT interval. J Am Coll Cardiol 2009;53(11):982-991. doi:10.1016/j.jacc.2008.12.014

21. Fourth Universal Definition of MI: with no LVH or LBBB, which new J-point ST elevation in V2-V3 meets the threshold in a woman?

  1. ≥ 0.5 mm
  2. ≥ 1.5 mm
  3. ≥ 2.5 mm
  4. ≥ 1 mm in a single lead
Show the answerHide the answer

Answer B: ≥ 1.5 mm. New J-point ST elevation in two contiguous leads suggests acute ischemia when it is ≥ 1 mm in leads other than V2-V3. In V2-V3 the cut-offs are ≥ 2 mm in men aged 40 or more, ≥ 2.5 mm in men under 40 and ≥ 1.5 mm in women of any age. These are ECG criteria only; diagnosing MI also needs the clinical picture and troponin.

  • A. ≥ 0.5 mm: 0.5 mm is too little in V2-V3; this definition uses 0.5 mm only for the extra posterior (V7-V9) and right-sided (V3R-V4R) leads.
  • B. ≥ 1.5 mm: Correct: ≥ 1.5 mm in V2-V3 for women, regardless of age.
  • C. ≥ 2.5 mm: 2.5 mm is the cut-off for men under 40 in V2-V3.
  • D. ≥ 1 mm in a single lead: The criteria need two contiguous leads, and V2-V3 use higher cut-offs than 1 mm.

References:

  1. Thygesen K, Alpert JS, Jaffe AS, et al. Fourth universal definition of myocardial infarction (2018). Circulation 2018;138(20):e618-e651. doi:10.1161/CIR.0000000000000617

22. Which situation calls for synchronized cardioversion rather than unsynchronized defibrillation?

  1. Ventricular fibrillation
  2. Pulseless ventricular tachycardia
  3. Asystole
  4. A regular tachycardia with a pulse causing hypotension and shock
Show the answerHide the answer

Answer D: A regular tachycardia with a pulse causing hypotension and shock. Synchronized cardioversion times the shock to the QRS, so it does not fall on the T wave, where it could trigger VF. The AHA uses it for an unstable tachyarrhythmia with a pulse. VF and pulseless VT are treated with unsynchronized defibrillation, and so is unstable polymorphic VT, because the device cannot reliably synchronize to it.

  • A. Ventricular fibrillation: VF has no organized QRS to synchronize to; it is treated with defibrillation.
  • B. Pulseless ventricular tachycardia: Pulseless VT is treated like VF, with unsynchronized defibrillation.
  • C. Asystole: Asystole is not a shockable rhythm.
  • D. A regular tachycardia with a pulse causing hypotension and shock: Correct: an unstable tachyarrhythmia with a pulse gets synchronized cardioversion.

References:

  1. American Heart Association. Adult tachyarrhythmia with a pulse algorithm, 2025. Source
  2. American Heart Association. Adult cardiac arrest algorithm, 2025. Source

23. A stable adult has a regular narrow-complex tachycardia, and vagal maneuvers have not worked. What does the AHA 2025 tachyarrhythmia algorithm suggest next?

  1. Atropine 1 mg IV
  2. Adenosine as a rapid IV push followed by a saline flush
  3. Unsynchronized defibrillation
  4. Adenosine is never used if vagal maneuvers fail
Show the answerHide the answer

Answer B: Adenosine as a rapid IV push followed by a saline flush. For a stable, regular narrow-complex tachycardia the algorithm suggests adenosine after vagal maneuvers. It is given as a rapid IV push with a saline flush, starting at 6 mg, with a larger second dose if needed. Adenosine briefly blocks the AV node; it may end a re-entrant SVT or reveal flutter waves.

  • A. Atropine 1 mg IV: Atropine speeds the heart; it is for symptomatic bradycardia.
  • B. Adenosine as a rapid IV push followed by a saline flush: Correct: adenosine, as a rapid IV push with a saline flush, is the next step here.
  • C. Unsynchronized defibrillation: Unsynchronized shocks are for VF and pulseless VT, not a stable rhythm with a pulse.
  • D. Adenosine is never used if vagal maneuvers fail: Adenosine is exactly the drug the algorithm suggests when vagal maneuvers fail.

References:

  1. American Heart Association. Adult tachyarrhythmia with a pulse algorithm, 2025. Source
  2. Page RL, Joglar JA, Caldwell MA, et al. 2015 ACC/AHA/HRS guideline for the management of adult patients with supraventricular tachycardia. Circulation 2016;133(14):e506-e574. doi:10.1161/CIR.0000000000000311

24. In the AHA 2025 adult bradycardia algorithm, what is the first drug and dose for unstable bradycardia?

  1. Atropine 0.5 mg IV, maximum 3 mg
  2. Adenosine 6 mg IV
  3. Atropine 1 mg IV, repeated every 3-5 min, maximum 3 mg
  4. Epinephrine 1 mg IV push
Show the answerHide the answer

Answer C: Atropine 1 mg IV, repeated every 3-5 min, maximum 3 mg. The current AHA algorithm gives atropine 1 mg IV, repeated every 3-5 minutes up to 3 mg, with pacing pads on. If atropine does not work, use transcutaneous pacing and/or an epinephrine or dopamine infusion. Atropine often fails when the block is below the AV node, such as complete block with a wide escape.

  • A. Atropine 0.5 mg IV, maximum 3 mg: 0.5 mg is an older dose; current AHA algorithms use 1 mg.
  • B. Adenosine 6 mg IV: Adenosine slows AV conduction; it is for regular narrow-complex tachycardia.
  • C. Atropine 1 mg IV, repeated every 3-5 min, maximum 3 mg: Correct: atropine 1 mg IV every 3-5 min, to a maximum of 3 mg.
  • D. Epinephrine 1 mg IV push: A 1 mg epinephrine push is the cardiac arrest dose; in bradycardia, epinephrine is a 2-10 mcg/min infusion.

References:

  1. Wigginton JG, Agarwal S, Bartos JA, et al. Part 9: adult advanced life support: 2025 American Heart Association guidelines for cardiopulmonary resuscitation and emergency cardiovascular care. Circulation 2025;152(16 Suppl 2):S538-S577. doi:10.1161/CIR.0000000000001376