Baseline Observe how the model values move over time when nothing is changed, and when one setting is changed.
Ventilation & CO2 Observe how PaCO₂ and pH follow a change in rate or tidal volume over the next minutes, and how much faster PaO₂ settles when ventilation is reduced.
ARDS & PEEP Observe how PaO₂ changes after a change in PEEP or FiO₂, and what Pplat and driving pressure do when PEEP changes.
Airflow obstruction & air trapping Observe whether expiratory flow returns to zero before the next breath, and how the intrinsic PEEP estimate changes with rate and I:E.
ARDS in pressure control Observe that in pressure control the tidal volume is a result: how VT, PaCO₂ and pH follow a change in inspiratory pressure, and how PaO₂ follows a change in PEEP.
Cases
Observation goal:
What this model does not show
Educational model, not clinical guidance. One model adult; every value is model output, not a measurement. The case covers up to 30:00 of patient time, and you can end it at any point.
Changes appear faster here than in patients. After each change the model fast-forwards so you see the result within a minute. In patients oxygenation settles over minutes and CO₂ over 10 to 30 minutes.
Model details
Start case Back to cases
Case 0:00
Left -
End case
Basic Advanced
Monitor Watch & record All records
Monitor model values
HR -- /min
SpO₂ -- %
MAP -- mmHg
EtCO₂ -- mmHg
ABP mmHg --
CO₂ mmHg -- EtCO₂
Mode Connecting
Adult PBW --
Breath --
Connecting… - Waiting for the engine Stop
No active alarms
Audio pause
Reset
Lung & loops
More values Values
Trend
Loops
-- mL/kg PBW Airway pressure -- cmH₂O
Per breath, last 40 breaths
Ppeak
VTe
EtCO₂
Inspiratory pause No qualified pause Measurement details
Waiting for engine observations.
Waiting for completed breaths
Settings just changed: breath values can still include breaths from before the change for about two breaths.
Waiting for timing Timing details in Model information
Model view · absolute lung volume
A model value, not a ventilator reading: the ventilator screen above shows tidal volume.
Lung volume L · absolute Fixed Now - L
Case complete
Copy text summary
Show the text summary
Your records, changes and predictions as plain text for study notes. Model output only; no names or dates.
What this model does not show
Your records
Swipe the table sideways to see every value.
Your predictions beside the model
Reference explanation Written from the engine's longer reference runs of this case, not from your run. Compare it with your own records above; a shorter run may show only the start of a change.
Download case log (JSON) Retry with a new session Back to cases
The case log holds model output only: case id and version, engine hash, methods, and the records and changes on the model clock.
All records and compare
Show all model values
Swipe the table sideways to see every value.
Compare two records
From To Compare
Blood gas (model) No samples Draw ABG ×
Results
Compare
Live
Blood gas results, one column per sample, newest first
No sample drawn yet.
Arrows show the change from the previous result. Model values, not laboratory results.
Continuous arterial values from the model. At the bedside you only see these after a drawn sample.
Live (model) Waiting for an observation
pH --
PaCO₂ mmHg --
PaO₂ mmHg --
HCO₃⁻ mmol/L --
Up to the last 20 minutes of simulation time. Dashed lines: setting changes. Dots: drawn samples.
Calculated by the model, not measured in a blood sample. How these are calculated
Draw ABG Blood gas · no samples
Guide ×
Start here This case
Changes appear faster here than in patients. After each change the model fast-forwards so you see the result within a minute. In patients oxygenation settles over minutes and CO₂ over 10 to 30 minutes.
How a case runs
Draw ABG (blood gas panel, right).
Change one setting on the tiles below the waveforms, then press Confirm.
Wait for the values to settle. With Fast-forward on, the case runs ahead and records the next blood gas by itself.
Draw ABG again if you want another sample.
Compare the results: Compare in the blood gas panel, or All records below.
Read the screen in this order
Monitor: HR, SpO₂, MAP, EtCO₂ (a separate device from the ventilator) Waveforms: Paw, Flow, Volume Measured values Blood gas: the case's signals to watch are marked ●
Blood gases here are instant model samples, calculated by the model, not collected blood samples. Locked settings show a lock and stay as the case set them.
More
Glossary Model information
No case running
Signals to watch
Locked in this case
Educational simulation · not for diagnosis. Connecting to the physiological engine…
Waiting for the next breath…
References Background reading on these topics. Values and observations in the cases are model output.
Acute Respiratory Distress Syndrome Network. Ventilation with lower tidal volumes as compared with traditional tidal volumes for acute lung injury and the acute respiratory distress syndrome. N Engl J Med 2000;342(18):1301-1308. doi:10.1056/NEJM200005043421801 Amato MBP, Meade MO, Slutsky AS, et al. Driving pressure and survival in the acute respiratory distress syndrome. N Engl J Med 2015;372(8):747-755. doi:10.1056/NEJMsa1410639 Grasselli G, Calfee CS, Camporota L, et al. ESICM guidelines on acute respiratory distress syndrome: definition, phenotyping and respiratory support strategies. Intensive Care Med 2023;49(7):727-759. doi:10.1007/s00134-023-07050-7 Hess DR. Respiratory mechanics in mechanically ventilated patients. Respir Care 2014;59(11):1773-1794. doi:10.4187/respcare.03410 Marini JJ. Dynamic hyperinflation and auto-positive end-expiratory pressure: lessons learned over 30 years. Am J Respir Crit Care Med 2011;184(7):756-762. doi:10.1164/rccm.201102-0226PP Tuxen DV, Lane S. The effects of ventilatory pattern on hyperinflation, airway pressures, and circulation in mechanical ventilation of patients with severe air-flow obstruction. Am Rev Respir Dis 1987;136(4):872-879. doi:10.1164/ajrccm/136.4.872 Leatherman JW, Ravenscraft SA. Low measured auto-positive end-expiratory pressure during mechanical ventilation of patients with severe asthma: hidden auto-positive end-expiratory pressure. Crit Care Med 1996;24(3):541-546. doi:10.1097/00003246-199603000-00028 Rittayamai N, Katsios CM, Beloncle F, et al. Pressure-controlled vs volume-controlled ventilation in acute respiratory failure: a physiology-based narrative and systematic review. Chest 2015;148(2):340-355. doi:10.1378/chest.14-3169
Waveform scale ×
Scale modeFixed Auto
Paw · cmH₂O−5 to 20 −5 to 40 −5 to 60 −5 to 80
Flow · L/min±60 ±90 ±120 ±180
Volume · mL0 to 400 0 to 600 0 to 800 0 to 1200
Fixed scales are the default, so a change in pressure, flow or volume stays visible. Auto fits all visible samples with headroom: it expands immediately and waits at least 5 seconds before shrinking. Flow is positive during inspiration and negative during expiration. The loops have a frame of their own: it starts at the fixed scale, grows at once when a loop would leave it and shrinks back after 5 seconds. The breath in progress is drawn as it happens, with a dot at the newest sample, over the previous three breaths in fading lines.
Unfreeze the display to change its scale.
Ventilation mode ×
Choose a mode and review its settings. Nothing changes until you confirm the mode.
Waiting for the engine's supported settings.
MV SIMULATOR
Alarm limits ×
Each bar shows a limit range with the current measured value as an arrow. Change a limit, or clear it to follow the settings again. Alarms are computed by this page from the measured values; the model does not raise them.
Auto-set limits Default limits
Alarm log model time
No alarms yet.
×
No engine connected.
Capabilities are checked before the session opens.
About this simulator
Physiology powered by Pulse Physiology Engine 4.3.2 , distributed under the Apache License 2.0 . These model observations are educational and are not clinical measurements.
Waveforms and blood values all come from one continuous simulated patient. What you set, what the model delivers and the blood values are shown as separate quantities.
The ventilator screen
The dark screen is a generic ventilator layout: mode and alarms on top, sweep waveforms on the left, measured values, Lung view and loops on the right, and settings at the bottom. Waveforms are drawn from the engine's original 50 Hz samples, left to right, with a small erased gap ahead of the newest sample; fixed scales are the default so changes stay visible. The volume waveform is tidal volume: lung volume above the end-expiratory lung volume of each breath.
Settings change like on a ventilator: select a tile, adjust it (drag, mouse wheel or arrow keys, Shift for bigger steps, or the slider on a phone), then Confirm or Cancel. An unconfirmed change goes back after 60 seconds. A new mode is reviewed in its own window and applied with Confirm mode. Notes about tidal volume per kg, PEEP plus inspiratory pressure, inverse I:E and short expiration never block a change; only the engine's own limits do.
What the numbers mean
Minute volume (MVe), respiratory rate, intrinsic PEEP (estimate) and end-tidal CO₂ are model ventilator outputs. Ppeak, measured PEEP, VTi, mean airway pressure over 6 breaths and measured I:E are read from the original samples of complete breaths, and so is VTe on the Simulator (in Cases, VTe is the model ventilator value). Pplat, driving pressure (DP), Cstat and Raw need a VC inspiratory pause and show -- without one. VT is the model's delivered tidal volume, separate from the set VT and from absolute lung volume (all the gas in the lungs), which is shown only in Model view. PBW is predicted body weight; for the model patient it equals the ideal body weight. Select any number for its definition and formula.
Arterial values are continuous model values, not collected blood samples. The bicarbonate basis is identified once the engine connects.
Alarms, Lung view and blood gases
Alarms are computed by this page from the measured values and the limits in Alarm limits; the model itself raises no alarms. High priority flashes red, medium flashes amber and low stays cyan, in the status bar only (no flashing when your system asks for reduced motion). Sound is off unless you turn on Alarm sound. SpO₂, heart rate and mean arterial pressure come from the model with every frame and are averaged like a bedside monitor (4, 8 and 2 seconds).
Lung view draws one large alveolus that inflates in place. Size: the air in the lung above residual volume. Residual volume: the air you cannot breathe out. Solid ring: the end of the last breath out; PEEP (plus any trapped air) sets it, and it stays while the next breath goes in. Dotted ring: one breath of 6 mL/kg PBW on top of the solid ring, so it moves with PEEP. Faint dashed ring: the model patient's healthy lung at rest without PEEP (FRC before any added lung condition), a reference and not a goal; a stiff or injured lung can end each breath out below it. The drawing stops at the patient's total lung capacity. Pink dashed ring while you adjust a VT or PEEP tile: a preview before Confirm. For VT, the end of the breath out plus the new VT (end of the breath in). For PEEP, the end of the breath out plus Cstat times the PEEP change, shown only after Cstat has been measured. It is an estimate that assumes compliance stays as last measured: in a real lung, recruitment can make the change larger and overdistension smaller. After Confirm it stays until one whole breath at the new setting has finished; then the solid ring (PEEP) or the alveolus at the end of each breath in (VT) shows the model's result. Color: the last plateau pressure, blue below 25 cmH₂O, turning amber from 25 to 30 and red from 30; gray until a pause (Tpause or an inspiratory hold) measures one, and again after a setting change. Pplat includes the chest wall. A driving pressure of 15 cmH₂O or more turns the DP label and the alveolus wall red. The color is a pressure cue only. One lung compartment: the view shows how much air the lung holds, not regional collapse or overstretch. The amber outline above 8 mL/kg PBW is a display threshold only: the model does not simulate overdistension injury. The arc shows airway pressure from 0 to 40 cmH₂O; the span from total PEEP to the last Pplat is the driving pressure. A patient without lung volumes shows the older picture instead: the alveolus fills with each breath relative to the reference tidal volume, and the dashed ring is 100 %.
On the Simulator page, Draw ABG copies the model's arterial values at that moment. As with a bedside blood gas analyzer, the result appears after 30 seconds of simulation time, in the Blood gas panel; a sample still in the analyzer when the session ends is not analyzed. The panel's Live tab shows the continuous model values, which at the bedside you only see through a drawn sample. P/F ratio, A-a gradient (PAO₂ = FiO₂ × (760 - 47) - PaCO₂ / 0.8, at sea level) and the PaCO₂ minus EtCO₂ gap are calculated from those values. A note marks a draw that may not be at steady state: PaCO₂ less than 20 minutes after a change in rate, VT, inspiratory pressure or mode, PaO₂ less than 5 minutes after a change in FiO₂ or PEEP. In a case, Draw ABG puts the values straight into the case's record table, with no analyzer wait, and the case shows the suggested wait after each change.
Freeze, Pause and the drawn line
Freeze holds the picture while the simulated patient keeps breathing. Pause stops simulation time. Both show the exact last sample. The line is drawn about 0.5 s behind the newest sample so it moves smoothly; every point is an original 50 Hz sample, and no values are filled in to hide a delay.
Fast-forward after each change: the model runs ahead without drawing until the values a change affects settle, for at most one minute of your time, so changes appear faster than in patients. One line above the monitor shows its progress and Stop. Blood gases are drawn only when you press Draw ABG; the Steady or Settling mark beside it says whether the arterial values are still moving. The box in the guide turns the fast-forward off.
What the model does not do
Modes are controlled (VC-CMV and PC-CMV): the model patient does not breathe spontaneously yet, so there are no triggered breaths.
Higher PEEP does not show overdistension harm in this model.
Higher PEEP raises the pressure around the heart: central venous pressure rises and cardiac output falls within seconds, and heart rate rises to hold blood pressure (default patient, PEEP 5 to 20: cardiac output about 11% lower, heart rate about 10 per minute higher, mean arterial pressure back to baseline within a minute). In a hypoxemic model patient, such as the ARDS case, heart rate can fall instead as oxygenation improves. Both heart rate patterns belong to this model; in patients stroke volume usually falls and heart rate holds. The effect is small when the lung is very stiff, pulse pressure variation does not follow PEEP yet, and the reflex holds blood pressure more fully than it would in a deeply sedated patient.
No kidney compensation: bicarbonate barely changes during a session.
No circuit leak, no sensor noise and no measured FiO₂ separate from the setting.
Reaching the high pressure alarm limit sounds the alarm but does not end the breath early, as many ventilators do.
No teaching exports, and no patient or hospital identity fields.
Technical details
Simulation time and elapsed real time appear once the session starts.
No accepted observation.
Ventilation terms ×
Plain-language meanings of the abbreviations on this page. They describe the terms; they are not settings advice.
Settings
VC · Volume Control A mode in which the ventilator delivers a set tidal volume with each breath. Airway pressure then rises to whatever level the resistance and compliance of the respiratory system require.
PC · Pressure Control A mode in which the ventilator holds airway pressure at a set level above PEEP for the inspiratory time. The delivered tidal volume and the flow pattern then depend on resistance and compliance.
VT · Tidal volume The volume of gas moved into the lungs in one breath, in mL. In VC it is a setting; the VT shown on the monitor is what the model actually delivered.
RR · Respiratory rate The number of breaths per minute (the Rate setting). Together with Ti or I:E it sets how much time each breath leaves for expiration.
PEEP · Positive end-expiratory pressure The airway pressure kept above atmospheric pressure at the end of expiration, in cmH₂O. It is the baseline of the pressure waveform between breaths.
FiO₂ · Fraction of inspired oxygen The share of oxygen in the delivered gas, shown as a percentage; room air is about 21%. It mainly changes oxygenation (PaO₂), not the shape of the waveforms.
Inspiratory pressure (PC) In Pressure Control, the pressure the ventilator adds above PEEP during inspiration, in cmH₂O.
Ti · Inspiratory time The time spent in inspiration in each breath, in seconds. Here it is the total inspiratory time: gas delivery plus any inspiratory pause.
I:E · Inspiratory-to-expiratory ratio The ratio of inspiratory time to expiratory time, written 1:x. A larger x leaves relatively more of each breath for expiration.
Tpause · Inspiratory pause A short period at the end of inspiration with no gas flow (VC only here). Pressure settles to a plateau during the pause, which is where Pplat is measured.
PC with volume target A pressure-controlled mode in which the ventilator adjusts the inspiratory pressure after each breath so the delivered tidal volume moves toward the set tidal volume, within a pressure limit. Ventilator makers use different names for this kind of mode.
Rise time In the pressure modes, the time the ventilator takes to bring airway pressure up to the set level at the start of inspiration, in seconds. A longer rise time gives a softer start and slightly less tidal volume.
Pressure limit The upper airway pressure of the volume-regulated pressure mode: the inspiratory pressure it chooses stays 5 cmH₂O below this limit, so a stiff lung may not reach the set volume.
Inspiratory flow pattern (VC) How gas flow is delivered during a Volume Control breath: constant (square) or a decelerating ramp that starts high and falls to zero. The same volume in the same time needs twice the starting flow with a ramp.
Waveforms and pressures
Paw · Airway pressure The pressure at the airway opening, in cmH₂O, shown in the top waveform.
Flow Gas flow at the airway, in L/min: positive during inspiration and negative during expiration. Expiratory flow that has not returned to zero when the next breath starts means the lungs had not finished emptying.
Volume · tidal The volume waveform on the ventilator screen: lung volume above the end-expiratory lung volume of the same breath, in mL. It starts near zero at each breath and peaks at about the tidal volume.
Lung volume · absolute All the gas in the lungs, in L, including the gas that stays in them between breaths. Tidal volume is the rise and fall of this line within each breath.
Ppeak · Peak inspiratory pressure The highest airway pressure during inspiration. It is the sum of PEEP, the pressure needed to drive flow through the airways and tube, and the pressure needed to stretch the respiratory system.
Pplat · Plateau pressure The airway pressure during an inspiratory pause, when flow is zero. Without flow there is no resistive pressure, so it reflects the elastic recoil of the lungs and chest wall at the end of inspiration, including PEEP.
DP · Driving pressure Plateau pressure minus total PEEP, in cmH₂O. It is the pressure that inflates the respiratory system for one tidal volume, so it needs an inspiratory pause to measure.
Pmean · Mean airway pressure The average airway pressure over whole breaths, in cmH₂O, here over the last 6 complete breaths.
Intrinsic PEEP · auto-PEEP Pressure left in the lungs at the end of expiration above the set PEEP, typically because expiration is too short for the lungs to empty fully. Set PEEP plus intrinsic PEEP is total PEEP; here it is a model estimate, not an expiratory-hold measurement.
Inspiratory hold A maneuver that stops all gas flow at the end of one inspiration for a set time. Airway pressure then settles to the plateau pressure, from which driving pressure, static compliance and (with a constant end-inspiratory flow) resistance are calculated.
ABP · Arterial blood pressure The pressure in the arteries over each heartbeat, in mmHg, shown as systolic/diastolic with the mean in brackets. Here it is the model value at 50 samples per second.
Capnogram CO₂ partial pressure at the Y-piece over each breath, in mmHg: near zero while fresh gas flows in, rising during expiration to the end-tidal value.
Calculated values
VTe and VTi · Expiratory and inspiratory tidal volume VTe is the volume breathed out in the last completed breath, in mL: on the Simulator the expiratory flow summed at the ventilator connection, in Cases the model ventilator value. VTi is the inspiratory flow of the last breath summed at the ventilator connection.
PBW · Predicted body weight A weight calculated from sex and height: male 50 + 0.91 × (height in cm - 152.4) kg, female 45.5 + 0.91 × (height in cm - 152.4) kg. Tidal volume is shown per kg of it (mL/kg PBW).
Minute ventilation The volume of gas moved in one minute, in L/min, about tidal volume × respiratory rate.
Resistance · inspiratory How strongly the airways and tube oppose gas flow, in cmH₂O·s/L. At the same flow, a higher resistance widens the gap between Ppeak and Pplat.
Crs · Respiratory-system compliance The volume the lungs and chest wall take up per unit of pressure, in mL/cmH₂O, estimated here as VT ÷ (Pplat − total PEEP). A lower compliance means a higher Pplat for the same VT.
Cstat · Static compliance The ventilator-screen name for respiratory-system compliance measured during an inspiratory pause, in mL/cmH₂O. It is the same quantity as Crs.
Raw · Airway resistance The ventilator-screen name for the inspiratory resistance estimate, in cmH₂O·s/L.
EtCO₂ · End-tidal CO₂ The CO₂ partial pressure at the end of an exhaled breath, in mmHg. It is usually somewhat lower than PaCO₂.
Arterial blood gases
pH The acidity of arterial blood. A rise in PaCO₂ lowers pH and a fall in PaCO₂ raises it.
PaO₂ · Arterial oxygen partial pressure The pressure of oxygen dissolved in arterial blood, in mmHg. It reflects how well oxygen moves from the lungs into the blood; FiO₂ is one of the settings that changes it.
PaCO₂ · Arterial carbon dioxide partial pressure The pressure of carbon dioxide in arterial blood, in mmHg. More alveolar ventilation lowers it and less ventilation raises it.
HCO₃⁻ · Bicarbonate The main buffer of the blood, in mmol/L. It changes only slightly when PaCO₂ changes quickly; larger changes take hours to days, so it moves little in a short session.
Inspiratory pause measurements ×
Waiting for a qualified pause.
In VC, Tpause is part of total inspiratory time. Increasing it shortens flow delivery and increases the flow required for the same set VT. Requested RR and expiration time are preserved; the native 50 Hz clock quantizes delivered timing.
Pplat is the mean pressure in a qualified, stable zero-flow portion of a completed inspiratory pause. The shaded interval marks the original samples used. Duration and stability affect the estimate; an unavailable value stays blank.
Inspiratory resistance ≈ (end-inspiratory pressure during constant flow − Pplat) / flow in L/s. This effective estimate includes the tube and other resistive components; it is not isolated airway resistance.
Ppeak · flow phase is the maximum pressure in the qualified constant-flow portion, excluding the initial transition. Resistance uses the terminal flow-phase pressure rather than a transient pressure spike.
Respiratory-system compliance (Crs) ≈ same-breath tidal volume / (Pplat − total PEEP). It includes lung and chest wall, not lung compliance alone. Total PEEP is a model estimate, not an expiratory-hold measurement.
A Pplat above 30 cmH₂O adds a short note under the value. In the ARDS Network trial of lower tidal volumes, the lower-volume group kept Pplat at or below 30 cmH₂O, measured after a 0.5 s end-inspiratory pause (N Engl J Med 2000;342:1301-8). The note is for teaching; it is not an alarm.
Results belong to the displayed breath and settings revision. Changing settings clears live results until another breath qualifies. Freeze preserves the original measurement and its time.