ECG rhythms Monitor pitfalls

Sinus rhythm with artifact ECG

Sinus rhythm with motion and muscle artifact. Narrow QRS complexes keep a regular rhythm under baseline wander and muscle noise.

Recorded stripHR —

A 30-second recording of the heart model, played in real time. Run live starts the engine itself with this rhythm.

Computed by the TracingsMD heart model (Junkers engine): simulated for education, not a patient recording and not for diagnosis.

Measured on this recording

From the engine's beat measurements over the last 10 seconds of the recording: rate from the mean RR interval, intervals as medians. Run live to watch them update.
Ventricular rate78/min
PRNot assessable in this rhythm
QRS84 ms
QTNot assessable in this rhythm
QTc (Bazett)Not assessable in this rhythm
QTc (Fridericia)Not assessable in this rhythm

What it is

Signal on the ECG that does not come from the heart: patient movement, shivering or muscle tremor, poor electrode contact, or electrical interference. Artifact can mimic ventricular fibrillation, ventricular tachycardia or atrial fibrillation, or, when a lead comes loose, asystole or a pause, while the real rhythm continues underneath. On this recording a regular sinus rhythm keeps its narrow QRS complexes under baseline wander and muscle noise.

How to read this strip

Rate60–100/min.
RhythmRegular QRS complexes under the noise.
P wavesPartly hidden by noise.
PRHard to measure through the noise.
QRSNarrow.
QTNot reliably measurable.

Common sources

What to consider

Key point

Before calling VF, look at the patient: if the patient is talking, find the QRS complexes marching through the noise.

Not to be confused with

Practice with this rhythm

References

  1. AHA practice standards for ECG monitoring in hospital settings (2017)
  2. AHA Adult Advanced Life Support (2025)